14.2 Log entries and error codes
Log entry
Parameter
Description
Code
Plain text
Code
Plain text
1020 ERROR SOD
##
---
Internal system error
There may be a hardware problem if this
error is contained in the device log several
times with the same parameter code.
Contact the manufacturer in this case (see
"Contact information" on page 111).
1031 UV_LOCKOUT
---
---
The device switches to energy saving mode
and terminates all of the operations as the
rechargeable battery or battery voltage is too
low. Only the charge controller, if present,
remains active.
1033 MODEM_UPDATE
##
---
Update of the firmware for the modem
controller was completed successfully.
This entry is always duplicated in the device
log. In the first entry, the parameter specifies
the major version number (e.g. 3 for
03v011), while in the second entry it
specifies the minor version number (e.g. 11
for 03v011).
1034 CONTROLLER_
UPDATE
##
---
Update of the firmware for the measurement
controller was completed successfully
This entry is always duplicated in the device
log. In the first entry, the parameter specifies
the major version number (e.g. 3 for
03v011), while in the second entry it
specifies the minor version number (e.g. 11
for 03v011).
1111
ALOHA STOP
0
---
End of ALOHA transmission mode
##
---
ALOHA transmission mode was stopped
unexpectedly. The parameter specifies how
many more seconds ALOHA transmission
mode should have lasted.
1200 MODEM ERROR
Modem error (see "Modem error" on page
98)
94
Rev. 01
Summary of Contents for myDatalogNANO Ex
Page 2: ......
Page 8: ......
Page 9: ...Chapter 2 Declaration of conformity Chapter 2 Declaration of conformity Rev 01 9 ...
Page 10: ...2 1 Declaration of conformity Appendix 10 Rev 01 ...
Page 11: ...Chapter 2 Declaration of conformity Rev 01 11 ...
Page 12: ......
Page 24: ......
Page 52: ......
Page 78: ......
Page 86: ......
Page 92: ......
Page 104: ......
Page 108: ......
Page 110: ......