7 T
ROUBLESHOOTING AND
M
AINTENANCE
See also:
l
Clean the Instrument on page 6 - 10
l
Instrument Status on page 1 - 5
l
Maintenance Menu Selection on page 6 - 1
l
Power Instrument On and Off on page 7 - 8
l
Routine Maintenance on page 7 - 9
l
Replace a Gas Cylinder on page 7 - 4
l
Printer Installation on page 6 - 6
Log in to your
to access error messages.
Sample tube rises when piston engages.
Cause:
The retaining sample tube collar is missing or too loose.
Action:
Fit the retaining collar tightly on the sample tube just below the bracket.
Unexpected test results.
Cause:
No sample, or incorrect sample in the tube.
Action:
Check that the correct sample has been loaded and the correct mass and density
have been entered.
Cause:
The top brass sample plug and filter paper are not in place.
Action:
Install top brass sample plug and filter paper.
Cause:
Sample has escaped the tube during the test.
Action:
Repeat the test using two filter papers on each plug.
Cause:
Inappropriate Fisher number mapping being applied
Action:
Check that the Fisher number map for the material is correct.
Sample escapes top of sample tube when piston first engages.
Cause:
The top O-ring is worn.
Action:
Check the integrity of the O-rings. See
.
Cause:
The sample material contains very fine particles.
Action:
The loss of a very small amount of sample may not require any action. A second filter
paper on the top brass plug may reduce the amount of sample lost.
7 Troubleshooting and Maintenance
MIC SAS II 5800 Operator Manual
580-42800-01 (Rev - ) — June 2019
7 - 1