S
AMPLE
S
ERVO
V
ALVE
Unit [n] > Calibration > Sample Servo
Unit [n] > Calibrate Analysis > Sample Servo
Use to calibrate the servo valve to the sample transducer. The servo valve should always be
recalibrated after a pressure calibration has been performed. The pressure transducer should be
calibrated before starting this calibration procedure.
Ensure the pressure transducer has been calibrated before performing this pro-
cedure. Go to
Unit [n] > Unit Configuration
and view the calibration information.
Contact your Micromeritics Service Representative if calibration dates are not listed.
Click
Start
. The window closes when the calibration is complete.
Click
Cancel
to stop the calibration process.
Sample Servo Valve
Gemini VII Operator Manual
239-42828-01 (Rev H ) — May 2021
9 - 5