S
PECIFICATIONS FOR THE
G
EMINI
VII
Surface Area
From 0.1 m
2
, total
From 0.01 m
2
/g, specific
Pore Volume
From 4 × 10
-6
cm
3
/g
Analysis
Frequency
50/60Hz
Power
50VA
Voltage
100-240VAC
Overvoltage Category
II
Electrical
Humidity
20% to 80% relative, non-condensing
Temperature
10 °C to 35 °C (50 °F to 95 °F), operating
0 °C to 50 °C (32 °F to 122 °F), non-operating
Indoor/Outdoor Use
Indoor only (not suitable for wet locations)
Altitude: 2000 m max
Pollution degree of the intended environment: 2
Degree of Ingress
Protection
IPX0
Environment
Adsorbate
Optimized for nitrogen in a liquid nitrogen sample bath. Gemini
may be used with non-corrosive adsorbate gases having vapor
pressures at both room and bath temperatures that are acceptably
high relative to the resolution of the 1,000 mmHg pressure trans-
ducer. Typically, oxygen, argon, carbon dioxide, butane, methane,
and other light hydrocarbons will produce useful data above abso-
lute pressures of a few mmHg.
Gases
Specifications for the Gemini VII
Gemini VII Keypad Operator Manual
239-42801-01 (Rev E) — July 2021
1 - 9