A D
ATA
F
ORMAT
Data can be transmitted in one of two user-selectable formats: single column or spreadsheet. The
data are in ASCII-delimited format. Spreadsheet format is suitable for direct import into many
popular spreadsheets.
In all tables, units are:
Date: DD/MM/YY
Time: HH:MM:SS
Pressure: mmHg Elapsed
Time: seconds
For all Pass/Fail reports:
0 = Fail
1 = Pass or Reported-only
R
EPORT
F
ORMAT
- S
INGLE
C
OLUMN
Record
Number
Information Conveyed
Form
1
Software version number
21 characters
2
Instrument ID
20 characters
3
Set Up group number
1 integer
4
Setup ID
20 characters
5
Sample ID
20 characters
6
Sample mass
1 floating point
7a
Start date
8 characters
7b
Start time
8 characters
8a
End date
8 characters
8b
End time
8 characters
9
Evacuation rate
1 floating point
10
Evacuation time
1 integer
11
Free space method:
0 = none
1 = previous
2 = measured
3 = calculated
1 integer
A Data Format
Gemini VII Keypad Operator Manual
239-42801-01 (Rev E) — July 2021
A - 1