3 Operate the Smart VacPrep
Field or Button
Description
Cancel
Discards any changes or cancels the current process.
Evacuate Options
l
Unrestricted evac. pressure.
Pressure value at which unrestricted
sample evacuation should begin.
l
Vacuum level.
Specify the vacuum level to be achieved before evac-
uation begins.
l
Evacuation rate.
Rate of evacuation.
Evacuate Port[s]
Select the ports() to evacuate.
Start
Starts the operation.
Smart VacPrep Evacuate Ports Fields and Buttons
Q
UICK
S
TART
D
EGAS
C
ONDITIONS
Smart VacPrep > Unit [n] > QuickStart Degas Conditions
The
QuickStart Degas Conditions
option provides the ability to enter commonly used degas conditions
for each Smart VacPrep port without the need for a sample information file. Degassing can then be
easily started using the front panel buttons on the Smart VacPrep unit.
This option provides three ways to load information into
QuickStart Degas Conditions
:
1. Load degas conditions from an existing degas conditions file.
2. Copy conditions from another Smart VacPrep unit and port.
3. Enter degas conditions manually.
Degas conditions can be copied into sample files after degassing for record keeping.
When using the
QuickStart
menu option, the degas process is started using the
Select
button and
Start
buttons for each port on the Smart VacPrep unit.
QuickStart
degas conditions are remembered for the next push button operation.
3 - 2
Smart VacPrep Operator Manual
Summary of Contents for Gemini VII 2390a
Page 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Page 4: ......
Page 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Page 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Page 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Page 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Page 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Page 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Page 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Page 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Page 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Page 264: ......
Page 266: ......
Page 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Page 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Page 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Page 380: ...Contact Us vi Smart VacPrep Operator Manual...
Page 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...