Appendix E
Gemini VII
E-4
239-42828-01 - Dec 2012
4.
Ensure that the filler rods are equidistant from the top of each tube. If they are not, add or remove
sample until they are the same distance from the top of the tube.
5.
Attach the sample tube to the analysis port and the balance tube to the balance port.
6.
Prepare your analysis Dewar and place it on the elevator.
7.
Close the sample compartment door and start your analysis.
Using Glass Beads and Filler Rods
This method may further improve results, depending on the sample material you are analyzing.
1.
After determining and loading the amount of glass beads to use, insert a filler rod into the balance
tube.
2.
Insert a filler rod into the sample tube. Ensure that the filler rods are equidistant from the top of
each tube. If they are not, add or remove sample until they are the same distance from the top of
the tube.
If you are using bulbous sample tubes, use the same size filler rod as used in the balance tube.
3.
Attach the sample tube to the analysis port and the balance tube to the balance port.
4.
Prepare your analysis Dewar and place it on the elevator.
5.
Close the sample compartment door and start your analysis.
Packing of some powders may restrict gas access to the powder and cause slower
equilibration and/or lower results.
Be sure to select Measure for the free-space measurement.
Summary of Contents for Gemini VII 2390a
Page 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Page 4: ......
Page 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Page 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Page 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Page 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Page 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Page 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Page 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Page 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Page 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Page 264: ......
Page 266: ......
Page 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Page 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Page 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Page 380: ...Contact Us vi Smart VacPrep Operator Manual...
Page 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...