E. ANALYZING SAMPLES WITH TOTAL SURFACE AREA OF
1.0M2 OR LESS
Introduction
The unique balanced measurement method used in the Gemini permits small amounts of surface area
to be measured with nitrogen gas that otherwise would be measurable only with krypton. Low surface
area samples often displace many times more nitrogen than they adsorb, especially if composed of
low-density materials of large particle size. The standard, built-in, helium free-space difference mea-
surement and mathematical compensation routine typically removes the effects of more than 99% of
this sample displacement, but the small amount remaining uncompensated can still be significant.
A technique for adding a compensating volume with negligible surface area into the balance tube has
been developed. It can reduce the size of the initial imbalance to low levels and allow precise measure-
ment of adsorbed gas.
This technique can be performed by the following methods:
• Using glass beads in the balance tube
• Using filler rods in the sample and balance tubes
Some users find that a combination of the two methods — using glass beads and filler rods — pro-
duces even better results. This, however, will depend on the type of sample you are analyzing.
Filler rods and glass beads are included in your accessories kit. Either straight-wall or bulb sample
tubes may be used.
Using Glass Beads
This method typically produces the best results.
1.
Place an appropriate quantity of sample in a clean sample tube.
v
w
----
=
2.
Load a second sample tube of the same size with glass beads that have a total volume
approximately the same as the sample volume.
a.) Determine the volume (
v
) of the sample in cm
3
:
where
w
= mass of sample (g)
= density of sample (g/cm
3
); if density is unknown, refer to your laboratory
handbook
Summary of Contents for Gemini VII 2390a
Page 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Page 4: ......
Page 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Page 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Page 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Page 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Page 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Page 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Page 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Page 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Page 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Page 264: ......
Page 266: ......
Page 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Page 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Page 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Page 380: ...Contact Us vi Smart VacPrep Operator Manual...
Page 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...