Sample Information Files
Gemini VII
5-8
239-42828-01 - Dec 2012
Sample Information Files
File > Open > Sample Information
(or use the
F2
keyboard shortcut)
Each analysis must be linked with a sample information file before the analysis can proceed. A sample
information file can consist of multiple parameter files; however, parameter files can also stand alone.
Parameter files allow repeated use of the file, for example, if the same analysis conditions exist for
multiple analyses, create an Analysis Conditions file containing the recurring conditions. When the
sample file is created, select the Analysis Conditions file for the analysis conditions. Once it becomes
part of the new sample file, edit the new file as needed without affecting the original Analysis Condi-
tions file. Sample Information files can be created or opened in Advanced, Basic, or Restricted
formats.
Specify or change the default format by selecting Options > Option Presentation or
select Basic / Advanced from the dropdown list at the bottom of the window. Refer to
for a description of the Advanced, Basic, and
Restricted formats
.
Field or Button
Description
Sample text box
Enter a sample description.
Operator / Submitter
text boxes
Enter the name of the operator or submitter performing the analysis.
Summary of Contents for Gemini VII 2390a
Page 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Page 4: ......
Page 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Page 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Page 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Page 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Page 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Page 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Page 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Page 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Page 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Page 264: ......
Page 266: ......
Page 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Page 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Page 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Page 380: ...Contact Us vi Smart VacPrep Operator Manual...
Page 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...