Installing the Vacuum Pump
Feb 09 Gemini VII 2390 Confirm Installation Instructions and Checklist
13
Installing the Exhaust Port Filter
1. If the exhaust port is closed with a clamp, loosen the wing nut of the clamp at the exhaust
port. Swing the clamp away from the exhaust port and remove it.
2. Make sure the centering ring remains on the port.
If the exhaust port is closed with a cap, remove the cap.
3. Place the filter on the centering ring. There are two types of centering rings, use the one
with the larger opening.
4. Open the clamp and place it around the flange of the exhaust port and the flange of the
exhaust filter.
5. Swing the clamp fastener screw toward the exhaust port until it fits into the slot in the
other half of the clamp. Tighten the wing nut securely by hand.