ASAP 2460 Operator’s Manual
Generating Graph Overlays
246-42800-01 - Aug 2013
2-53
Generating Graph Overlays
Use the graph overlay function to compare multiple graph options. Graphical lines are differentiated
by the use of varying colored symbols outlined on a legend. Overlays may be generated in two ways:
•
Multiple Sample Overlays -
overlay up to 25 plots of the same type with that of the current plot.
•
Multiple Graph Overlays -
overlay two different types of plots from one sample. This type of
overlay is available only for:
– BJH Adsorption/Desorption
– DFT Pore Size/Surface Energy
– Dollimore-Heal Adsorption/Desorption
– Horvath-Kawazoe
– M-P Method
Only the Advanced format can be used to generate overlays. Go to
Options > Options Presentation >
Advanced
to access the Advanced format or select
Advanced
from the dropdown list at the bottom of
the window.
Multiple Sample Overlays
To overlay the same type of graph on multiple samples:
1.
Go to
File
>
Open
.
2.
Select the .SMP file and click
Open
.
If a file with a status other than
Preparing
,
Prepared
, or
No Analysis
is selected, the isotherm plot
displays. Select
Advanced
from the dropdown list at the bottom of the window.
When working with an existing file, a copy of the file should be used rather than the
original.
Summary of Contents for ASAP 2460
Page 4: ......
Page 12: ...Table of Contents ASAP 2460 Operator s Manual viii 246 42800 01 Nov 2013...
Page 30: ...Using the Software ASAP 2460 Operator s Manual 1 18 246 42800 01 Aug 2013...
Page 93: ...ASAP 2460 Operator s Manual Generating Graph Overlays 246 42800 01 Aug 2013 2 63...
Page 94: ...Generating Graph Overlays ASAP 2460 Operator s Manual 2 64 246 42800 01 Aug 2013...
Page 202: ...Service Test ASAP 2460 Operator s Manual 4 36 246 42800 01 Aug 2013...
Page 230: ...Report Examples 2460 Operator s Manual 5 28 246 42800 01 Aug 2013 Isotherm Linear Plot...
Page 231: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 29 BET Surface Area Report...
Page 232: ...Report Examples 2460 Operator s Manual 5 30 246 42800 01 Aug 2013 BET Surface Area Plot...
Page 233: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 31 t Plot Report...
Page 236: ...Report Examples 2460 Operator s Manual 5 34 246 42800 01 Aug 2013...
Page 242: ...Service Test Mode 2460 Operator s Manual 6 6 246 42800 01 Aug 2013...
Page 270: ...Calibrating the System ASAP 2460 Operator s Manual 7 28 246 42800 01 Aug 2013...
Page 276: ...Ordering Information ASAP 2460 Operator s Manual 8 6 246 42800 01 Aug 2013...
Page 278: ......
Page 280: ......
Page 374: ...Appendix C ASAP 2460 Operator s Manual C 50 246 42800 01 Aug 2013...
Page 398: ...Appendix F ASAP 2460 Operator s Manual F 18 246 42800 01 Aug 2013...
Page 412: ...Index ASAP 2460 Operator s Manual Index 8 246 42800 01 Aug 2013...
Page 413: ...ASAP 2460 Operator s Manual Appendix G Aug 2013 SVP 1 SVP Smart VacPrep Operator Manual...
Page 414: ...Appendix G ASAP 2460 Operator s Manual SVP 2 Aug 2013...
Page 422: ...Contact Us vi Smart VacPrep Operator Manual...
Page 432: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 466: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...