8 E
RROR
M
ESSAGES
Program error messages are listed numerically. If the
Action
response indicates to contact a
Micromeritics service representative, record the error message, then make backup copies of any files
involved in the operation.
4407
Error searching for installed Smart VacPreps. The registry could not be read..
Cause:
The application was not installed properly.
Action:
Reinstall the application. If the problem persists, contact your Micromeritics service rep-
resentative.
4408
The %s in %s already has Smart VacPrep S/N %d. The Smart VacPrep
must be removed from the %s before it can be added.
Cause:
The Smart VacPrep was already installed for another application / unit.
Action:
Remove the Smart VacPrep from the installed unit before adding it to the preferred unit.
4409
A free IP address on the same subnet as %s could not be found."
Cause:
All IP addresses on the network for the Ethernet card specified during installation are in
use by other Micromeritics applications on this computer.
Action:
Uninstall unused Micromeritics applications.
Action:
Configure a different Ethernet card for use by the application using the application
installer.
4410
The .INI file could not be updated with configuration for Smart
VacPrep S/N %s."
Cause:
The application .INI file is opened by another application and could not be updated.
Action:
Close all open applications and add the Smart VacPrep again using the Smart VacPrep
menu.
4411
Error dosing.
Cause:
The backfill timed out.
Action:
Ensure there is gas available and the pressure regulator is set to the appropriate pressure.
Also ensure that the gas supply regulator shut-off valve is open.
8 Error Messages
Smart VacPrep Operator Manual
8 - 1
Summary of Contents for ASAP 2460
Page 4: ......
Page 12: ...Table of Contents ASAP 2460 Operator s Manual viii 246 42800 01 Nov 2013...
Page 30: ...Using the Software ASAP 2460 Operator s Manual 1 18 246 42800 01 Aug 2013...
Page 93: ...ASAP 2460 Operator s Manual Generating Graph Overlays 246 42800 01 Aug 2013 2 63...
Page 94: ...Generating Graph Overlays ASAP 2460 Operator s Manual 2 64 246 42800 01 Aug 2013...
Page 202: ...Service Test ASAP 2460 Operator s Manual 4 36 246 42800 01 Aug 2013...
Page 230: ...Report Examples 2460 Operator s Manual 5 28 246 42800 01 Aug 2013 Isotherm Linear Plot...
Page 231: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 29 BET Surface Area Report...
Page 232: ...Report Examples 2460 Operator s Manual 5 30 246 42800 01 Aug 2013 BET Surface Area Plot...
Page 233: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 31 t Plot Report...
Page 236: ...Report Examples 2460 Operator s Manual 5 34 246 42800 01 Aug 2013...
Page 242: ...Service Test Mode 2460 Operator s Manual 6 6 246 42800 01 Aug 2013...
Page 270: ...Calibrating the System ASAP 2460 Operator s Manual 7 28 246 42800 01 Aug 2013...
Page 276: ...Ordering Information ASAP 2460 Operator s Manual 8 6 246 42800 01 Aug 2013...
Page 278: ......
Page 280: ......
Page 374: ...Appendix C ASAP 2460 Operator s Manual C 50 246 42800 01 Aug 2013...
Page 398: ...Appendix F ASAP 2460 Operator s Manual F 18 246 42800 01 Aug 2013...
Page 412: ...Index ASAP 2460 Operator s Manual Index 8 246 42800 01 Aug 2013...
Page 413: ...ASAP 2460 Operator s Manual Appendix G Aug 2013 SVP 1 SVP Smart VacPrep Operator Manual...
Page 414: ...Appendix G ASAP 2460 Operator s Manual SVP 2 Aug 2013...
Page 422: ...Contact Us vi Smart VacPrep Operator Manual...
Page 432: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 466: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...