Schematic Shortcut Icon Available Options:
o
Cool the sample to less than 50 °C
o
Set heater power percent
Smart VacPrep Schematic Shortcuts Table (continued)
S
HOW
S
MART
V
AC
P
REP
S
TATUS
Smart VacPrep > Unit [n] > Show Status
The
Smart VacPrep Status
window allows the monitoring of degas operations for a manual degas
operation.
Field or Button
Description
Cancel
Cancels the current degas operation.
Check
Click to check the outgassing rate of the samples on the selected port.
The following process occurs:
l
The current degassing step is suspended.
l
The vacuum valves are closed and the vacuum level on the selected
ports is monitored.
l
The
Status
window indicates that the degassing operation is being
checked and displays the outgassing rate as it becomes available.
During the degas check, the
Check
button changes to
Continue
. When
Continue
is clicked, the valves open and the degassing operation
resumes.
If the outgassing rate indicated that the sample has been freed of
contaminants (minimal pressure increases), click
Skip
to advance to the
next state of the degassing operation. For example, if degassing is
checked after the setpoint is attained,
Skip
advances you to the ramping
stage.
Smart VacPrep Status Fields and Buttons Table
Show Smart VacPrep Status
Smart VacPrep Operator Manual
3 - 11
Summary of Contents for ASAP 2460
Page 4: ......
Page 12: ...Table of Contents ASAP 2460 Operator s Manual viii 246 42800 01 Nov 2013...
Page 30: ...Using the Software ASAP 2460 Operator s Manual 1 18 246 42800 01 Aug 2013...
Page 93: ...ASAP 2460 Operator s Manual Generating Graph Overlays 246 42800 01 Aug 2013 2 63...
Page 94: ...Generating Graph Overlays ASAP 2460 Operator s Manual 2 64 246 42800 01 Aug 2013...
Page 202: ...Service Test ASAP 2460 Operator s Manual 4 36 246 42800 01 Aug 2013...
Page 230: ...Report Examples 2460 Operator s Manual 5 28 246 42800 01 Aug 2013 Isotherm Linear Plot...
Page 231: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 29 BET Surface Area Report...
Page 232: ...Report Examples 2460 Operator s Manual 5 30 246 42800 01 Aug 2013 BET Surface Area Plot...
Page 233: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 31 t Plot Report...
Page 236: ...Report Examples 2460 Operator s Manual 5 34 246 42800 01 Aug 2013...
Page 242: ...Service Test Mode 2460 Operator s Manual 6 6 246 42800 01 Aug 2013...
Page 270: ...Calibrating the System ASAP 2460 Operator s Manual 7 28 246 42800 01 Aug 2013...
Page 276: ...Ordering Information ASAP 2460 Operator s Manual 8 6 246 42800 01 Aug 2013...
Page 278: ......
Page 280: ......
Page 374: ...Appendix C ASAP 2460 Operator s Manual C 50 246 42800 01 Aug 2013...
Page 398: ...Appendix F ASAP 2460 Operator s Manual F 18 246 42800 01 Aug 2013...
Page 412: ...Index ASAP 2460 Operator s Manual Index 8 246 42800 01 Aug 2013...
Page 413: ...ASAP 2460 Operator s Manual Appendix G Aug 2013 SVP 1 SVP Smart VacPrep Operator Manual...
Page 414: ...Appendix G ASAP 2460 Operator s Manual SVP 2 Aug 2013...
Page 422: ...Contact Us vi Smart VacPrep Operator Manual...
Page 432: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 466: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...