Degas ports can also be canceled using buttons on the Smart VacPrep
unit. Press
Select
on the unit to select the port, then press the
Cancel
but-
ton.
S
HOW
S
MART
V
AC
P
REP
D
ASHBOARD
Smart VacPrep > Unit [n[ > Show Dashboard
The dashboard displays the following:
l
Number of days until roughing pump maintenance is due
l
Manifold outgas rate
Data for the dashboard comes from the logged diagnostic data. The dashboard is automatically kept
current as the relevant diagnostic data items are updated. The gauges will be updated even if the
dashboard window is not open.
Red numbers on the dashboard require attention. To reset the dashboard
numbers, right click on the dashboard setting, then click
Reset
.
Field or Button
Description
Days until roughing-pump
service is due
Annual maintenance is recommended. The number of days until the
anniversary of the last pump maintenance is shown. The displayed value
is updated at least once per day and when the maintenance time is reset.
When the displayed value is 30 or less, the value is displayed in red. Red
negative numbers display if maintenance is past due.
Manifold outgas rate
Provides the qualitative indication of the outgas rate in the dosing
manifold. LED images constitute a bidirectional bar graph of the outgas
rate.
The gauge is updated after each outgas rate measurement.
Smart VacPrep Dashboard Gauges and Descriptions Table
Show Smart VacPrep Dashboard
Smart VacPrep Operator Manual
3 - 7
Summary of Contents for ASAP 2460
Page 4: ......
Page 12: ...Table of Contents ASAP 2460 Operator s Manual viii 246 42800 01 Nov 2013...
Page 30: ...Using the Software ASAP 2460 Operator s Manual 1 18 246 42800 01 Aug 2013...
Page 93: ...ASAP 2460 Operator s Manual Generating Graph Overlays 246 42800 01 Aug 2013 2 63...
Page 94: ...Generating Graph Overlays ASAP 2460 Operator s Manual 2 64 246 42800 01 Aug 2013...
Page 202: ...Service Test ASAP 2460 Operator s Manual 4 36 246 42800 01 Aug 2013...
Page 230: ...Report Examples 2460 Operator s Manual 5 28 246 42800 01 Aug 2013 Isotherm Linear Plot...
Page 231: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 29 BET Surface Area Report...
Page 232: ...Report Examples 2460 Operator s Manual 5 30 246 42800 01 Aug 2013 BET Surface Area Plot...
Page 233: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 31 t Plot Report...
Page 236: ...Report Examples 2460 Operator s Manual 5 34 246 42800 01 Aug 2013...
Page 242: ...Service Test Mode 2460 Operator s Manual 6 6 246 42800 01 Aug 2013...
Page 270: ...Calibrating the System ASAP 2460 Operator s Manual 7 28 246 42800 01 Aug 2013...
Page 276: ...Ordering Information ASAP 2460 Operator s Manual 8 6 246 42800 01 Aug 2013...
Page 278: ......
Page 280: ......
Page 374: ...Appendix C ASAP 2460 Operator s Manual C 50 246 42800 01 Aug 2013...
Page 398: ...Appendix F ASAP 2460 Operator s Manual F 18 246 42800 01 Aug 2013...
Page 412: ...Index ASAP 2460 Operator s Manual Index 8 246 42800 01 Aug 2013...
Page 413: ...ASAP 2460 Operator s Manual Appendix G Aug 2013 SVP 1 SVP Smart VacPrep Operator Manual...
Page 414: ...Appendix G ASAP 2460 Operator s Manual SVP 2 Aug 2013...
Page 422: ...Contact Us vi Smart VacPrep Operator Manual...
Page 432: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 466: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...