Field or Button
Description
Save
Saves changes to the active window.
Smart VacPrep Evac-
uation
This option is applicable only when degassing with a Smart VacPrep.
l
Backfill sample tube.
Indicate if the sample tube should be backfilled
automatically or wait for operator response.
l
Evacuation Rate.
Rate used for evacuation.
l
Unrestricted evac. from.
Pressure at which the unrestricted evac-
uation is to begin.
l
Vacuum level.
Pressure for unrestricted evacuation.
l
Evacuation time.
Length of time for preliminary evacuation before
proceeding with the
Heating Phase
temperature schedule. The timer
starts when the vacuum level is reached.
l
Temperature ramp rate.
Rate at which the temperature is to change
when advancing to the target pressure.
l
Target temperature.
Targeted pressure for evacuation.
l
Hold pressure.
Pressure at which heating will stop and hold the
sample temperature approximately constant until the pressure falls
below the
Hold
pressure. This prevents damage to the sample struc-
ture due to 'steaming' and /or elutriation due to excessive escaping gas
velocity.
Degas Conditions Fields and Buttons Table (continued)
S
TART
S
MART
V
AC
P
REP
D
EGAS
Smart VacPrep > Unit [n] > Start Degas
This option allows the degassing of up to six samples simultaneously. A sample can be added to any idle
port and a degas operation started. When adding a sample, the degas operation in progress will be
suspended until the sample reaches the target pressure, then degassing on all ports will resume. If other
samples are still in the initial evacuation phase, the new samples will wait until the fast evacuation is
started. Samples can be removed from any of the six ports without disturbing the degas operations
already in process. Starting several samples as a batch allows them to perform the initial evacuation at
the same time.
Start Smart VacPrep Degas
Smart VacPrep Operator Manual
3 - 5
Summary of Contents for ASAP 2460
Page 4: ......
Page 12: ...Table of Contents ASAP 2460 Operator s Manual viii 246 42800 01 Nov 2013...
Page 30: ...Using the Software ASAP 2460 Operator s Manual 1 18 246 42800 01 Aug 2013...
Page 93: ...ASAP 2460 Operator s Manual Generating Graph Overlays 246 42800 01 Aug 2013 2 63...
Page 94: ...Generating Graph Overlays ASAP 2460 Operator s Manual 2 64 246 42800 01 Aug 2013...
Page 202: ...Service Test ASAP 2460 Operator s Manual 4 36 246 42800 01 Aug 2013...
Page 230: ...Report Examples 2460 Operator s Manual 5 28 246 42800 01 Aug 2013 Isotherm Linear Plot...
Page 231: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 29 BET Surface Area Report...
Page 232: ...Report Examples 2460 Operator s Manual 5 30 246 42800 01 Aug 2013 BET Surface Area Plot...
Page 233: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 31 t Plot Report...
Page 236: ...Report Examples 2460 Operator s Manual 5 34 246 42800 01 Aug 2013...
Page 242: ...Service Test Mode 2460 Operator s Manual 6 6 246 42800 01 Aug 2013...
Page 270: ...Calibrating the System ASAP 2460 Operator s Manual 7 28 246 42800 01 Aug 2013...
Page 276: ...Ordering Information ASAP 2460 Operator s Manual 8 6 246 42800 01 Aug 2013...
Page 278: ......
Page 280: ......
Page 374: ...Appendix C ASAP 2460 Operator s Manual C 50 246 42800 01 Aug 2013...
Page 398: ...Appendix F ASAP 2460 Operator s Manual F 18 246 42800 01 Aug 2013...
Page 412: ...Index ASAP 2460 Operator s Manual Index 8 246 42800 01 Aug 2013...
Page 413: ...ASAP 2460 Operator s Manual Appendix G Aug 2013 SVP 1 SVP Smart VacPrep Operator Manual...
Page 414: ...Appendix G ASAP 2460 Operator s Manual SVP 2 Aug 2013...
Page 422: ...Contact Us vi Smart VacPrep Operator Manual...
Page 432: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 466: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...