Appendix F
ASAP 2460 Operator’s Manual
F-14
246-42800-01 - Aug 2013
quently leads to inconsistencies when carried to small mesopore sizes. If the thickness curve used is
too steep, it finally will predict a larger increment of adsorptive for a given pressure increment than is
actually observed; since a negative pore volume is non-physical, the algorithm must stop. Conversely,
if the thickness curve used underestimates film thinning, accumulated error results in the calculation of
an overly large volume of (possibly nonexistent) small pores.
The use of equation (1) represents an improvement over the traditional algorithm. Kernel functions
corresponding to various classical Kelvin-based methods have been calculated for differing geometries
and included in the list of models.
Models Included
t
3.54
5.00
–
P
/Po
ln
----------------------
1/3
=
Kelvin Equation with Halsey Thickness Curve
N2 - Halsey Thickness Curve
The kernel function is calculated using the Halsey equation with standard parameters:
The nitrogen properties used in the Kelvin equation are:
N2 - Halsey Thickness Curve
The calculation is the same as above except that cylindrical geometry is assumed.
Geometry
Slit
Substrate
Average
Category:
Porosity
Method:
Nitrogen at 77 K
Surface tension =
8.88 dynes cm
-1
Molar density =
0.02887 g cm
-3
Geometry
Cylinder
Substrate
Average
Category:
Porosity
Method:
Nitrogen at 77 K
Reference
:
G. Halsey, J. Chem. Phys 16, 931 (1948).
Summary of Contents for ASAP 2460
Page 4: ......
Page 12: ...Table of Contents ASAP 2460 Operator s Manual viii 246 42800 01 Nov 2013...
Page 30: ...Using the Software ASAP 2460 Operator s Manual 1 18 246 42800 01 Aug 2013...
Page 93: ...ASAP 2460 Operator s Manual Generating Graph Overlays 246 42800 01 Aug 2013 2 63...
Page 94: ...Generating Graph Overlays ASAP 2460 Operator s Manual 2 64 246 42800 01 Aug 2013...
Page 202: ...Service Test ASAP 2460 Operator s Manual 4 36 246 42800 01 Aug 2013...
Page 230: ...Report Examples 2460 Operator s Manual 5 28 246 42800 01 Aug 2013 Isotherm Linear Plot...
Page 231: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 29 BET Surface Area Report...
Page 232: ...Report Examples 2460 Operator s Manual 5 30 246 42800 01 Aug 2013 BET Surface Area Plot...
Page 233: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 31 t Plot Report...
Page 236: ...Report Examples 2460 Operator s Manual 5 34 246 42800 01 Aug 2013...
Page 242: ...Service Test Mode 2460 Operator s Manual 6 6 246 42800 01 Aug 2013...
Page 270: ...Calibrating the System ASAP 2460 Operator s Manual 7 28 246 42800 01 Aug 2013...
Page 276: ...Ordering Information ASAP 2460 Operator s Manual 8 6 246 42800 01 Aug 2013...
Page 278: ......
Page 280: ......
Page 374: ...Appendix C ASAP 2460 Operator s Manual C 50 246 42800 01 Aug 2013...
Page 398: ...Appendix F ASAP 2460 Operator s Manual F 18 246 42800 01 Aug 2013...
Page 412: ...Index ASAP 2460 Operator s Manual Index 8 246 42800 01 Aug 2013...
Page 413: ...ASAP 2460 Operator s Manual Appendix G Aug 2013 SVP 1 SVP Smart VacPrep Operator Manual...
Page 414: ...Appendix G ASAP 2460 Operator s Manual SVP 2 Aug 2013...
Page 422: ...Contact Us vi Smart VacPrep Operator Manual...
Page 432: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 466: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...