E. MAINTAINING HIGH PURITY GASES
The system was designed to accurately measure the surface area of all types of materials. It is impor-
tant that the gases (especially krypton) used for these measurements be of highest purity, especially
when analyzing low surface area samples. Three ways to ensure high purity gases are to always
maintain:
• thoroughly purged gas pressure regulators
• non-permeable gas lines
• leak-free connections
Impure gas is strongly indicated, for example, if a series of measurements on a low surface area mate-
rial yields decreasing specific surface areas with decreasing quantities of sample. The system uses very
small amounts of helium; therefore any residual air in the regulator can distort results of subsequent
analyses for quite some time.
Micromeritics offers the following suggestions to assist you in maintaining high purity gases (particu-
larly helium) for the instrument.
• Use metal gas lines only
• Remove trapped air from the regulator and gas lines
Using Metal Gas Lines
You should always use metal gas lines which have been carefully cleaned of any oils and greases used
in the manufacturing process.
Do not use plastic or rubber gas lines
. When these types of permeable,
nonmetallic gas lines are used with helium, contaminants accumulate at a much faster rate. This causes
errors in analysis results and can also contaminate a clean sample.
Removing Trapped Air
When connecting the regulator to the gas bottle, air is unavoidably trapped on the high- and low-pres-
sure sides of the regulator, as well as in the gas lines. You should remove as much of this air as is
possible
before
opening the gas bottle valve. If this air is allowed to remain in the regulator, it will mix
with the helium and cause inaccurate results in subsequent analyses. Or if the valve is open for any
length of time, the air trapped on the high pressure side may diffuse back into the gas bottle and con-
taminate its entire contents.
There are two methods for removing trapped air from the regulator lines: the Purge Method and the
Evacuation Method.
Summary of Contents for ASAP 2460
Page 4: ......
Page 12: ...Table of Contents ASAP 2460 Operator s Manual viii 246 42800 01 Nov 2013...
Page 30: ...Using the Software ASAP 2460 Operator s Manual 1 18 246 42800 01 Aug 2013...
Page 93: ...ASAP 2460 Operator s Manual Generating Graph Overlays 246 42800 01 Aug 2013 2 63...
Page 94: ...Generating Graph Overlays ASAP 2460 Operator s Manual 2 64 246 42800 01 Aug 2013...
Page 202: ...Service Test ASAP 2460 Operator s Manual 4 36 246 42800 01 Aug 2013...
Page 230: ...Report Examples 2460 Operator s Manual 5 28 246 42800 01 Aug 2013 Isotherm Linear Plot...
Page 231: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 29 BET Surface Area Report...
Page 232: ...Report Examples 2460 Operator s Manual 5 30 246 42800 01 Aug 2013 BET Surface Area Plot...
Page 233: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 31 t Plot Report...
Page 236: ...Report Examples 2460 Operator s Manual 5 34 246 42800 01 Aug 2013...
Page 242: ...Service Test Mode 2460 Operator s Manual 6 6 246 42800 01 Aug 2013...
Page 270: ...Calibrating the System ASAP 2460 Operator s Manual 7 28 246 42800 01 Aug 2013...
Page 276: ...Ordering Information ASAP 2460 Operator s Manual 8 6 246 42800 01 Aug 2013...
Page 278: ......
Page 280: ......
Page 374: ...Appendix C ASAP 2460 Operator s Manual C 50 246 42800 01 Aug 2013...
Page 398: ...Appendix F ASAP 2460 Operator s Manual F 18 246 42800 01 Aug 2013...
Page 412: ...Index ASAP 2460 Operator s Manual Index 8 246 42800 01 Aug 2013...
Page 413: ...ASAP 2460 Operator s Manual Appendix G Aug 2013 SVP 1 SVP Smart VacPrep Operator Manual...
Page 414: ...Appendix G ASAP 2460 Operator s Manual SVP 2 Aug 2013...
Page 422: ...Contact Us vi Smart VacPrep Operator Manual...
Page 432: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 466: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...