ASAP 2460 Operator’s Manual
Appendix D
246-42800-01 - Aug 2013
D-1
D. FREE SPACE CORRECTION
Free space is that volume of the sample tube which is unoccupied by the sample. The quantity of gas
dosed into the sample tube is calculated from the difference in pressures in the manifold before and
after the dose is delivered. The quantity of gas adsorbed by the sample is calculated by subtracting the
quantity of gas remaining in the free space of the sample tube after equilibrium is established from the
quantity of gas originally dosed into the sample tube. Free space must be determined accurately to
obtain a precise value for quantity adsorbed.
Static-volumetric systems consist basically of a gas manifold joined to a sample tube by an isolation
valve. The manifold section has connections for an absolute pressure transducer, a temperature gauge,
and a vacuum system. It also has inlets for the adsorptive gas and helium. A dewar flask containing a
cryogenic liquid (usually LN
2
at approximately 77 K) is situated so that it can be raised to immerse
most of the sample tube. Two temperature zones exist within the sample tube when immersed in the
cryogenic bath: a warm zone (the volume above the liquid level and near ambient temperature) and a
cold zone (the volume below the liquid level at cryogenic temperature). Not only must the total free
space volume be determined, but it also is necessary to determine the quantity of gas residing within
the “cold” zone since a nonideality correction must be applied to only that quantity of gas.
The total quantity of gas in the partly immersed sample holder cannot simply be determined using n =
PV/RT because temperature is not constant over the total volume, but instead is distributed as two tem-
perature zones with a steep temperature gradient between them. A convenient method for resolving
this problem is to derive two factors which, for the existing temperature profile, can be multiplied by
the prevailing pressure to reveal the molar volume of gas contained in the cold zone and the total quan-
tity residing in the free volume of the immersed sample holder (the cold free space).
The system provides the following methods for free space determination:
• Measure
• Calculate
• Enter
Measure
Generally, this method, although requiring a little more time (approximately 10 minutes), is the most
preferred one for determining free space. It is simple, automatic, requires very little information, and
essentially is error-proof. With this method, the instrument first evacuates the manifold and sample
tube (containing sample), then isolates the sample tube by closing the valve. Then the manifold is
charged with helium, the pressure measured, and the valve opened allowing the helium to expand into
the sample tube at ambient temperature. Again the pressure is measured.
The dewar is raised and the sample tube is cooled to cryogenic temperature. Again pressure drops;
when pressure has equilibrated, the value is recorded. Warm and cold free spaces are calculated from
(1) system volume, (2) system, ambient, and bath temperatures, and (3) measured pressures. From
these, the value of the portion of cold free space at cryogenic temperature which requires correction for
nonideality can be determined.
This method may be undesirable if:
Summary of Contents for ASAP 2460
Page 4: ......
Page 12: ...Table of Contents ASAP 2460 Operator s Manual viii 246 42800 01 Nov 2013...
Page 30: ...Using the Software ASAP 2460 Operator s Manual 1 18 246 42800 01 Aug 2013...
Page 93: ...ASAP 2460 Operator s Manual Generating Graph Overlays 246 42800 01 Aug 2013 2 63...
Page 94: ...Generating Graph Overlays ASAP 2460 Operator s Manual 2 64 246 42800 01 Aug 2013...
Page 202: ...Service Test ASAP 2460 Operator s Manual 4 36 246 42800 01 Aug 2013...
Page 230: ...Report Examples 2460 Operator s Manual 5 28 246 42800 01 Aug 2013 Isotherm Linear Plot...
Page 231: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 29 BET Surface Area Report...
Page 232: ...Report Examples 2460 Operator s Manual 5 30 246 42800 01 Aug 2013 BET Surface Area Plot...
Page 233: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 31 t Plot Report...
Page 236: ...Report Examples 2460 Operator s Manual 5 34 246 42800 01 Aug 2013...
Page 242: ...Service Test Mode 2460 Operator s Manual 6 6 246 42800 01 Aug 2013...
Page 270: ...Calibrating the System ASAP 2460 Operator s Manual 7 28 246 42800 01 Aug 2013...
Page 276: ...Ordering Information ASAP 2460 Operator s Manual 8 6 246 42800 01 Aug 2013...
Page 278: ......
Page 280: ......
Page 374: ...Appendix C ASAP 2460 Operator s Manual C 50 246 42800 01 Aug 2013...
Page 398: ...Appendix F ASAP 2460 Operator s Manual F 18 246 42800 01 Aug 2013...
Page 412: ...Index ASAP 2460 Operator s Manual Index 8 246 42800 01 Aug 2013...
Page 413: ...ASAP 2460 Operator s Manual Appendix G Aug 2013 SVP 1 SVP Smart VacPrep Operator Manual...
Page 414: ...Appendix G ASAP 2460 Operator s Manual SVP 2 Aug 2013...
Page 422: ...Contact Us vi Smart VacPrep Operator Manual...
Page 432: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 466: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...