Analysis Conditions Files
ASAP 2460 Operator’s Manual
3-28
246-42800-01 - Aug 2013
Equilibration
button
Click to enter values for relative or absolute pressure.
•
Relative Pressure (p/p
o
)
or
Absolute Pressure (mmHg) -
the pressure
the equilibration interval will be applied.
•
Equilibration Interval -
the number of seconds between successive
pressure readings during equilibration. At least 11 pressure readings for
each equilibration are required to obtain an accurate reading of the
pressure changes.
Long equilibration intervals tend to lengthen analyses, but improve data
integrity.
Short equilibration intervals produce a faster analysis but may reduce
the accuracy of data.
•
Minimum equilibration delay at p/p
o
> =0.995
- the minimum number
of seconds required before equilibration can occur for a relative pressure
greater than or equal to 0.995. This field is not available if
Absolute
pressure dosing
is selected on the
Analysis
Conditions
tab.
Field or Button
Description (continued)
Column heading
reads Absolute
Pressure (mmHg)
when Absolute
Pressure Dosing is
selected
Does not
display when
Absolute
Pressure
Dosing is
selected.
Summary of Contents for ASAP 2460
Page 4: ......
Page 12: ...Table of Contents ASAP 2460 Operator s Manual viii 246 42800 01 Nov 2013...
Page 30: ...Using the Software ASAP 2460 Operator s Manual 1 18 246 42800 01 Aug 2013...
Page 93: ...ASAP 2460 Operator s Manual Generating Graph Overlays 246 42800 01 Aug 2013 2 63...
Page 94: ...Generating Graph Overlays ASAP 2460 Operator s Manual 2 64 246 42800 01 Aug 2013...
Page 202: ...Service Test ASAP 2460 Operator s Manual 4 36 246 42800 01 Aug 2013...
Page 230: ...Report Examples 2460 Operator s Manual 5 28 246 42800 01 Aug 2013 Isotherm Linear Plot...
Page 231: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 29 BET Surface Area Report...
Page 232: ...Report Examples 2460 Operator s Manual 5 30 246 42800 01 Aug 2013 BET Surface Area Plot...
Page 233: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 31 t Plot Report...
Page 236: ...Report Examples 2460 Operator s Manual 5 34 246 42800 01 Aug 2013...
Page 242: ...Service Test Mode 2460 Operator s Manual 6 6 246 42800 01 Aug 2013...
Page 270: ...Calibrating the System ASAP 2460 Operator s Manual 7 28 246 42800 01 Aug 2013...
Page 276: ...Ordering Information ASAP 2460 Operator s Manual 8 6 246 42800 01 Aug 2013...
Page 278: ......
Page 280: ......
Page 374: ...Appendix C ASAP 2460 Operator s Manual C 50 246 42800 01 Aug 2013...
Page 398: ...Appendix F ASAP 2460 Operator s Manual F 18 246 42800 01 Aug 2013...
Page 412: ...Index ASAP 2460 Operator s Manual Index 8 246 42800 01 Aug 2013...
Page 413: ...ASAP 2460 Operator s Manual Appendix G Aug 2013 SVP 1 SVP Smart VacPrep Operator Manual...
Page 414: ...Appendix G ASAP 2460 Operator s Manual SVP 2 Aug 2013...
Page 422: ...Contact Us vi Smart VacPrep Operator Manual...
Page 432: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 466: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...