ASAP 2460 Operator’s Manual
Analysis Conditions Files
246-42800-01 - Aug 2013
3-23
Preparation
button
•
Fast evacuation
- Select
Fast evacuation
when analyzing samples (such
as pellets) that do not fluidize or shed particles under evacuation.
•
Evacuation rate / Unrestricted evac. from -
enter the evacuation rate
and the pressure at which unrestricted sample evacuation is to begin. The
option is enabled when Fast evacuation is not selected.
•
Evacuation time -
enter the length of time for preliminary evacuation
which takes place prior to the free space measurement or sample analysis
if free space is to be Entered or Calculated.
•
Leak test -
enables the system to check for system leaks or sample
outgassing before the analysis.
The leak test seals the sample tube and monitors the pressure. If the
pressure rises more than 0.15 mmHg, the analysis does not proceed and a
message displays. While leak testing slightly increases analysis time, it
prevents the continuation of analysis and collection of erroneous data if a
leak exists.
Select to enable the
Leak test duration
field. Enter how long the pressure
is to be monitored.
•
Use TranSeal
- select if using the TranSeal to transfer the sample from
the preparation port to the analysis port under vacuum.
Field or Button
Description (continued)
Summary of Contents for ASAP 2460
Page 4: ......
Page 12: ...Table of Contents ASAP 2460 Operator s Manual viii 246 42800 01 Nov 2013...
Page 30: ...Using the Software ASAP 2460 Operator s Manual 1 18 246 42800 01 Aug 2013...
Page 93: ...ASAP 2460 Operator s Manual Generating Graph Overlays 246 42800 01 Aug 2013 2 63...
Page 94: ...Generating Graph Overlays ASAP 2460 Operator s Manual 2 64 246 42800 01 Aug 2013...
Page 202: ...Service Test ASAP 2460 Operator s Manual 4 36 246 42800 01 Aug 2013...
Page 230: ...Report Examples 2460 Operator s Manual 5 28 246 42800 01 Aug 2013 Isotherm Linear Plot...
Page 231: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 29 BET Surface Area Report...
Page 232: ...Report Examples 2460 Operator s Manual 5 30 246 42800 01 Aug 2013 BET Surface Area Plot...
Page 233: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 31 t Plot Report...
Page 236: ...Report Examples 2460 Operator s Manual 5 34 246 42800 01 Aug 2013...
Page 242: ...Service Test Mode 2460 Operator s Manual 6 6 246 42800 01 Aug 2013...
Page 270: ...Calibrating the System ASAP 2460 Operator s Manual 7 28 246 42800 01 Aug 2013...
Page 276: ...Ordering Information ASAP 2460 Operator s Manual 8 6 246 42800 01 Aug 2013...
Page 278: ......
Page 280: ......
Page 374: ...Appendix C ASAP 2460 Operator s Manual C 50 246 42800 01 Aug 2013...
Page 398: ...Appendix F ASAP 2460 Operator s Manual F 18 246 42800 01 Aug 2013...
Page 412: ...Index ASAP 2460 Operator s Manual Index 8 246 42800 01 Aug 2013...
Page 413: ...ASAP 2460 Operator s Manual Appendix G Aug 2013 SVP 1 SVP Smart VacPrep Operator Manual...
Page 414: ...Appendix G ASAP 2460 Operator s Manual SVP 2 Aug 2013...
Page 422: ...Contact Us vi Smart VacPrep Operator Manual...
Page 432: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 466: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...