Starting an Analysis
AccuPyc II 1340
4-4
June 2013
Starting an Analysis
Before you start your analysis, be sure you have:
• specified analysis and report parameters, refer to pages
• prepared and loaded your sample, page
The prompts that display for an analysis depend on the options you choose in
Set Up >
Analysis Parameters
and
Report Options
; therefore, your analysis may not display all of the
prompts referenced in this procedure.
1.
Press
Alt + 4
; the
Sample ID
prompt is shown; enter an appropriate identification. You
can enter up to 20 numbers and dashes; press
.
(decimal) to insert a dash.
2.
Press
ENTER
; the
Description
prompt is displayed. You can enter up to 20 numbers and
dashes on each line; press
.
(decimal) to insert a dash.
Most users display this prompt only when a computer keyboard is attached to the
pycnometer.
3.
Press
ENTER
; the
Sample mass
prompt is displayed if you have chosen
Density
at the
Anls display mode
prompt in
Set Up > Report Options
. Enter the mass (weight) of your
sample.
If you have an analytical balance connected to the RS232 port, press the appropriate
button on the balance to transfer the sample mass while this prompt is displayed.
4.
Press
ENTER
; the
Chamber insert
prompt is displayed if you are using an insert. Press
CHOICE
until the appropriate insert is displayed.
5.
Press
ENTER
to choose the insert.
6.
Press
ENTER
to begin the analysis; operational status messages display during analysis.
7.
When the analysis has finished, the pycnometer beeps three times and the display returns
to the
Reload
prompt.
8.
Be sure the pressure reads approximately 0 (zero) on the display, then remove the cell
chamber cap and remove the sample from the chamber.
9.
Replace the cell chamber cap, or load another sample.
Do not remove the cell chamber cap when the pycnometer is pressurized. Sample may
be discharged from the chamber.
Summary of Contents for AccuPyc 2 1340
Page 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Page 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Page 4: ......
Page 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Page 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Page 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Page 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Page 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Page 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Page 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Page 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Page 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Page 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Page 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Page 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...