4. GENERAL OPERATING INSTRUCTIONS
This chapter contains step-by-step procedures for common operations for the AccuPyc.
Preparing for Analysis
Setting Regulator Pressure
The Purge fill pressure and Cycle fill pressure are specified using
Set Up > Analysis Parame-
ters
). Set the regulator pressure at the higher of the two (plus 2.0
psig). For example; if you have specified the Purge fill pressure as
19.500
psig and the Cycle
fill pressure as
19.000
psig, you should set your regulator pressure at
21.500
psig.
1.
Be sure the tank pressure for the gas regulator is at least 200 psig. Pressures less than 200
psig indicate the tank is low on gas. Analyses are terminated automatically if gas is
depleted.
2.
Press
Alt + 1
to enter manual mode.
3.
Press
8
(EXPAND) and
9
(VENT) to open the expansion and vent valves; the second line
of the display should look like this:
X - O - O
4.
Press
7
(FILL) to open the fill valve; the second line of the display should look like this:
O - O - O
5.
Adjust the regulator pressure control knob until the desired pressure is shown on the
regulator display.
6.
Press
7
to close the fill valve, then increase the regulator valve by 2.0 psig; the second
line of the display should look like this:
X - O - O
.
7.
Allow the pressure in the pycnometer to drop below 2.0 psig, then press
8
and
9
to close
the Expansion and Vent valves; the display should look like this:
X - X - X
8.
Press
SAVE
to return to the
Reload
prompt.
Pressure Control Knob
Summary of Contents for AccuPyc 2 1340
Page 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Page 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Page 4: ......
Page 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Page 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Page 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Page 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Page 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Page 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Page 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Page 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Page 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Page 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Page 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Page 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...