AccuPyc II 1340
System Commands
Aug 2013
3-23
Analyze
This command enables you to perform an analysis; press
Alt
+
4
on the keypad.
At any time during the following sequence of prompts, you can:
• Press
SAVE
to store the data you entered and return to the
Reload
prompt; no analysis
is started. This also deletes any previously stored sample data. Verify that all desired
reports from the previous sample have been viewed before pressing
SAVE.
• Press
Alt
+
CLEAR
to discard the information you entered and return to the
Reload
prompt.
The prompts that display during the
Analyze
function are dependent upon the choices you make in the
Setup > Report Options
function explained on page
The date is included in all printed and transmitted reports,
and is set before shipment. However, the date shown in the
display may be incorrect for the area to which the instrument
is shipped.
The date is entered in a two-digit format for the day, month,
and year. For example; to set the date for September 1, 2006,
enter
01/09/06
(day/month/year). Use the
.
(decimal) key to
enter a slash (/).
Press
ENTER
to advance to the next prompt.
The time (in 24-hour format) at which an analysis begins
and ends is shown on all printed and transmitted reports. The
time shown on the display may be inappropriate for your
time zone and, therefore, may require changing.
The time is entered in a two-digit format for the hour, min-
utes, and seconds. For example; to set the time to 1:05 pm,
enter
13:05:00
(hour:minutes:seconds). Use the
.
(decimal)
key to enter a colon (
:
).
Press
ENTER
to advance to the next prompt.
Unit[n]
SN1234 10 cm3
Unit Types
Date (DD/MM/YY)?
(pre-set date)
Unit[n]
SN1234 10 cm3
Unit Types
Time (HH:MM:SS)?
(pre-set time)
Summary of Contents for AccuPyc 2 1340
Page 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Page 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Page 4: ......
Page 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Page 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Page 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Page 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Page 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Page 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Page 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Page 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Page 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Page 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Page 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Page 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...