AccuPyc II 1340
Verifying Operation
June 2013
2-15
Verifying Operation
This section provides instructions for performing an analysis on the calibration standard(s) to
verify operation of your AccuPyc. You do not have to edit set-up values; the same values used
in the calibration will be used in the analysis.
1.
Press
Alt + 4
; a prompt similar to the following is displayed:
2.
Press
ENTER
to begin the analysis. When the analysis has finished, the system
automatically returns to the
Reload
prompt.
3.
Press
CHOICE
to view the average volume and standard deviation.
4.
Compare the value displayed for the average volume to the value recorded on the top of
your calibration kit. These two values should agree plus or minus the variance obtained
using the following calculation:
Variance = (chamber volume * 0.0003) + (Value recorded on Standards case * 0.0003)
If the values do not agree, a leak is indicated. Refer to
Checking the Cell and Expansion
and test the system for leaks. Then repeat steps 1 through
3. If the values continue to disagree, contact your service representative.
5.
Press
Alt + 6
to print a hard copy of the report for your records.
6.
Remove the standard(s) and place them back into the case.
It is recommended that you read Chapter 3,
User Interface
before performing analyses. This
will familiarize you with the hardware and software of the AccuPyc.
Example
: calculate the variance for a
10-cm
3
pycnometer with a value of
6.372242
recorded on the case as follows:
Variance
= (10 * 0.0003) + (6.372242 * 0.0003) = 0.003 + 0.0019116726 =
0.005
Unit[n]
SN1234
10 cm3
Analyze
[ENTER] to start
[ESCAPE] to cancel
Unit[n]
SN1234
10 cm3
X - X - X
Avg Vol = (volume)
Std Dev = (deviation)
Summary of Contents for AccuPyc 2 1340
Page 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Page 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Page 4: ......
Page 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Page 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Page 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Page 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Page 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Page 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Page 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Page 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Page 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Page 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Page 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Page 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...