Installing the Analyzer
June 2013
2-3
Installing the Analyzer
The remainder of this chapter describes how to install the pycnometer and verify that it is
operating properly.
Location
When selecting the location of the pycnometer, keep the following in mind:
• The pycnometer performs best in a constant temperature environment.
• It should be installed on a workbench about 36 in. (91.4 cm) high in a location free of
drafts from either a forced-air heating or cooling system.
• The pycnometer should not be placed near a window; exposure to sunlight may cause
the temperature to vary.
Gas Requirements
The pycnometer uses helium or nitrogen (99.995% pure or better) to provide rapid, accurate
analyses. The cylinder containing helium must be fitted with a gas regulator set for 21-23 psig
(145-159 kPag). The pressure input to the pycnometer should never be greater than 25 psig
(172 kPag). Excessive pressures waste gas due to a protection device contained in the pycnom-
eter that vents the pycnometer to atmospheric pressure if pressure exceeds 35 psig (241 kPag).
This protection device is not installed on 2000 cm
3
units..
Other inert, dry gases can also be used and may be more applicable for some applications.
Minimum recommended gas tank pressure is 200 psig (1379 kPag).
Summary of Contents for AccuPyc 2 1340
Page 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Page 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Page 4: ......
Page 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Page 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Page 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Page 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Page 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Page 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Page 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Page 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Page 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Page 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Page 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Page 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...