Appendix F
AccuPyc II 1340
F-2
Nov 06
Adding an Analysis Module
Use these instructions when adding an Analysis module to an existing Temperature-controlled
unit.
1. Be sure the controlling unit is turned off.
2. Connect the circulating bath as instructed in the previous section.
3. Attach one end of the module connector cable (provided in the analysis module
accessories kit) to the connector labeled
Module Connection
on the rear panel of the
single-unit system.
4. Attach the other end of the module connector cable to the lower connector labeled Module
Connector on the rear panel of the analysis module
Be sure to use the lower connector on the analysis module; the upper connector is for
attaching another analysis module.
5. Place the On/Off switch on the rear panel of the analyzer in the On ( | ) position.
6. Calibrate the Analysis module using the Calibration Kit shipped with your system. Refer
to
Calibrating the Pycnometer
in Chapter 2 for step-by-step instructions.
If you have the Windows version of the AccuPyc, exit the application before turning off
the analyzer.
Use these ports to
connect the bath.
Summary of Contents for AccuPyc 2 1340
Page 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Page 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Page 4: ......
Page 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Page 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Page 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Page 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Page 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Page 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Page 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Page 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Page 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Page 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Page 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Page 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...