Run Precision
Run precision requires at least five runs. Run precision criterion is met when the sample vol-
umes calculated for the four most recent previous runs fall within the specified error band for
the current run’s sample volume. The error band is a specified percentage of the nominal vol-
ume of the sample chamber (1, 10, 100, or 350 cm
3
and 2000 cm
3
).
V
samp
[0] to
V
samp
[4] are the five most recent sample volumes,
V
samp
[4] being the most recent.
where
V
nominal
= nominal sample cell volume
If
for
i
= 0 to 3, run precision is achieved.
Total Pore Volume
where
D
bulk
= entered bulk density
Total Solids Concentration
where
D
liq
= entered liquid density
D
sol
= entered solid density
ErrBand
PctFullScale
V
nominal
100
----------------
=
V
samp
i
V
samp
4
–
ErrBand
TotalPoreVol
D
samp
D
bulk
–
D
samp
D
bulk
------------------------------
=
WeightPctSolids
1
D
liq
D
samp
------------
D
sol
D
sol
D
liq
–
-----------------------
100
–
=
Summary of Contents for AccuPyc 2 1340
Page 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Page 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Page 4: ......
Page 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Page 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Page 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Page 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Page 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Page 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Page 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Page 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Page 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Page 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Page 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Page 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...