Appendix A
AccuPyc II 1340
A-12
Oct 2013
ZEROERR: Eq failure
ZEROERR: Overrange
ZEROERR: Underrange
A.
Cause:
The system failed to equilibrate within 1000 seconds.
Action:
Check the pycnometer for leaks. Make sure the pycnometer has been
placed in a draft-free environment.
Cause:
A pressure overrange occurred while zeroing.
Action:
Check for a fill valve leak. If there is none, contact a Micromeritics
Service Representative.
Cause:
An error occurred in the pressure measurement electronics while
zeroing, causing a pressure underrange to occur.
Action:
Call a Micromeritics Service Representative.
Summary of Contents for AccuPyc 2 1340
Page 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Page 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Page 4: ......
Page 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Page 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Page 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Page 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Page 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Page 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Page 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Page 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Page 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Page 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Page 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Page 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...