This predictive clinical digital electronic thermometer is for the intermittent  

measurement and monitoring of human adult and pediatric body temperatures; 

orally, and under the arm. 

This thermometer can be used with or without 

a tip cover. To use a cover, slide onto sensor 

tip until fully engaged and peel off paper.  

Covers are for single use, do not reuse. If 

 thermometer has been in a cold environment, 

allow it to warm up to room temperature for 

about 15 minutes before use. Avoid storing at  

temperatures above 140

˚

 (60

˚

 C). 

This thermometer IS intended to be used: 

• For self-measurement/monitoring of human body temperature whereas the  

person being measured may be the user/operator of the device. 

• Orally, or Under the Arm (axillary). 

• Adults 

• Pediatric

 

This thermometer IS NOT intended for use with: 

• Neonatal 

 

General Warnings 

• Contact your physician in case of prolonged fever as this requires medical  

attention, especially for young children. 

• Self-measuring means monitoring, not diagnosis or treatment. Under no  

circumstances should you independently alter the dosages of any drugs  

prescribed by your physician. 

• Do not use this device if you think it is damaged or anything appears unusual 

(e.g. sporadic operation, open packaging upon purchase). 

• To avoid possible injury, do not run or walk while using this device. 

• Ensure that children are supervised around this medical device; some parts are 

small enough to be swallowed. 

• Keep batteries away from children; harmful if swallowed. 

• Do not dispose of the battery or device in fire; the battery may explode. 

• Do not use a metal tool to replace the battery; doing so may short circuit the terminals. 

• Do not attempt to recharge the battery as they are not rechargeable.  

• Only open the thermometer battery cover to replace the battery. 

• Do not open/modify this device; inaccuracy and/or harmful injury may result. 

• Performance of this thermometer may be degraded if used outside of its specified 

operation temperature and humidity rating, or if stored outside of its specified 

storage temperature and humidity rating. Avoid dust or storage in direct sunlight. 

• Performance of this thermometer may be degraded if the device is exposed to  

mechanical shock. 

• Performance of this thermometer may be degraded if the patient’s temperature is 

lower than that of the ambient. 

• Performance of this thermometer may be degraded if the tip is dirty or damaged; 

ensure the tip is clean before using. 

• Performance of this thermometer may be degraded if the probe cover is dirty or 

a probe cover not intended for use with this device is used. 

• Any probe covers that may be included in this device package are for single use only. 

• This device contains sensitive electronics components. Avoid strong electrical or 

electromagnetic fields in the direct vicinity of the device (e.g., mobile telephones,  

microwave ovens). These can lead to temporary impairment of the measuring  

accuracy. Move the device to another location if interference is determined. 

Maintain a minimum distance of 3.3 m from such devices when using this unit. 

• According to international standards, your device should be checked for  

accuracy every 2 years. 

How to Take a Temperature: 

Press power button to turn unit on. Display should read “188.8” to confirm  

normal function. An “Err” message indicates unit is not working properly. The 

last recorded measurement will show for 2 seconds, followed by “- - -” and a 

flashing “F” or “C”, indicating that thermometer is ready for use. 

Oral Use: 

• Do not drink hot or cold liquids or exercise before use. 

•  Place sensor tip under tongue as shown in Figure A. 

• Accurate temperature reading will take approximately 30 

seconds. An open mouth or improper placement could 

make reading take longer.

 

Underarm Use: 

• Wipe underarm with dry towel. 

• Place sensor tip in armpit and lower arm against body as 

shown in Figure B. 

• Accurate temperature reading will take approximately 5 

minutes with proper use.

 

Reading the Temperature: 

For readings 99.5

˚

 F (37.5

˚

 C) and below, the thermometer 

beeps 10 times to indicate temperature is within normal range.  

For readings above 99.5

˚

 F (37.5

˚

 C), the thermometer emits a series of short, 

rapid beats to indicate a possible high temperature. 

 

 
 
 
 

 

Changing from °F to °C: 

With the thermometer turned off, press and hold the power button for 5  

seconds. An “°F” or “°C” will appear on the display, blinking. Press the 

ON/OFF button to toggle to the opposite temperature setting. After 4 seconds 

your thermometer will be ready for use. 

How to Clean: 

Clean thermometer after every use. There are 2 cleaning options below: 

1. Alcohol: 

Clean

 sensor tip with 70% isopropyl alcohol. Wipe dry with 

sterile cotton ball.  

2. Dishwasher Safe: To clean in dishwasher, place thermometer on top rack only. 

Run dishwasher on normal cleaning cycle. High heat drying cycle is NOT recom-

mended. 

 

How to Change Battery: 

Replace battery when “ ” appears on display. Replace old battery with new 

battery (1.5V, LR41 or equivalent). Plus (+) sign should face upward. Replace 

endcap to close. 

 

Limited Warranty: 

This thermometer is warrantied for 5 years by Microlife USA Inc, against  

manufacturer defects for the original purchaser only, from date of purchase. 

The 5 year warranty applies to the thermometer. Batteries and case are not 

covered by this warranty. There are no user serviceable parts inside.  

The warranty does not apply to consequential and incidental damages, or 

damage caused by batteries, improper handling, and accidents. Professional 

use, not following the operating instructions, and alterations made to the  

device or accessory by third parties, are also not included in this warranty. 

Some states do not allow the exclusion or limitation of incidental or consequential 

damages, so the above limitation or exclusion may not apply to you. 

Microlife USA Inc will investigate your concern. A product as defined by this 

warranty, determined to be out of specification, will be replaced and shipped 

to you at no cost. A product or accessory as defined by this warranty, determined 

to be within specification, will be returned to you with a report of findings, at 

no cost. 

Please call customer service: 1-800-568-4147 regarding any warranty concerns. 

We ask that you please contact us before sending any product back in order 

to better identify, and more quickly process, your concern. 

This warranty gives you specific legal rights, and you may also have other 

rights which vary from state to state. 

 

Technical Specifications: 

Measurement range: 

90.0° F to 109.9° F  (32.0°C to 43.9° C) 

Measurement Accuracy: 

±0.2° F (±0.1° C) between 93.2° F (34° C) to 

107.6° F (42° C)  

Self test: 

Automatic internal check of sensor circuit. If 

error, «Err» is displayed.  

Display: 

Liquid crystal display (LCD) with four digits 

Auto Off:

With reading: after 9 minutes of inactivity  

Without reading: after 1 minute of inactivity 

Smallest Unit of Display:

0.1° F (0.1° C) 

Memory: 

For storing the last measured value 

Operating Temperature: 

50° F(10 °C) to 104° F (40° C) humidity of 15% to 

95% (non-condensing) 

Storage temperature:

 -13° F (-25° C) to 140° F (60° C) humidity of 15% 

to 95% (non-condensing) 

Battery: 

1.5V, LR41 

Expected product life:

5 years 

Disposal:

Batteries and electronic instruments must be 

disposed of in accordance with the locally ap-

plicable regulations, not with domestic waste. 

Reference to Standards: 

Device Standard: Device corresponds to the 

requirements of the standard for electronic 

thermometers ASTM E-1112, ISO 80601-2-56, 

IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11 as 

well as  ISO 10993-10:2010, ISO 10993-5: 2009, 

ISO 10993-12:2012 

IP67:

Totally protected against dust ingress,  

protected against the effect of immersion  

between 15cm and 1m. 

Electromagnetic Compatibility: Device fulfills the stipulations of the standard 

IEC 60601-1-2 

Test Protocol (for laboratory use only): 

1. Place thermometer into waterbath. The tip of the thermometer should be ap-

proximately 2cm below the surface. 

2. Wait for 10 seconds, then turn the thermometer on by pressing the “On/Off” 

button. The thermometer is now in standard mode. 

3. When the measurement is completed, record the temperature reading. 

Distributed by: 

Microlife USA, Inc. 

1617 Gulf to Bay Blvd., 2nd Floor 

Clearwater, FL 33755 

©2020 Microlife Corporation 
www.microlifeusa.com

30-Second 
Dishwasher Safe Thermometer 

Model #MT1P21-PRO

Sensor Tip

Flexible Probe

LCD Display

Power Button

(Figure B)

(Figure A)

Follow instructions of use to avoid hazards.

Type BF applied part

MT1P21-PRO_0820_4    

Made in China

Questions? 

We’re here to help. Call toll-free 

1-800-568-4147

Function Check

Last Recorded 

Temperature

Ready to 

Measure

End Cap