5
Microlife Ear Thermometer IR 310
EN
Recall mode
AQ
:
Press the M-button
3
to enter recall mode
when the power is off. The memory icon
«M»
will show.
Reading 1 - the last reading
AR
:
Press and release the M-
button
3
to recall the last reading. Number «1» and a flashing
«M»
are displayed.
Pressing and releasing the M-button
3
after the last 30 readings
have been recalled will resume the above sequence from reading 1.
9. Error Messages
Measured temperature too high
AS
:
Displays
«H»
when
measured temperature is higher than 43 °C / 109.4 °F.
Measured temperature too low
BT
:
Displays
«L»
when
measured temperature is lower than 32 °C / 89.6 °F.
Ambient temperature too high
BK
: Displays
«AH»
when
ambient temperature is higher than 40.0 °C / 104.0 °F.
Ambient temperature too low
BL
: Display
«AL»
when
ambient temperature is lower than 10.0 °C / 50.0 °F.
Incorrect location indication
AL
:
Display
«ERP
» when the
probe is not correctly inserted in the ear canal. Please insert the
probe as described in this manual.
No probe cover on the measuring sensor
AM
:
Displays
probe cover icon
AM
with a red LCD backlight to remind the
user to put on a new undamaged probe cover before each
measurement.
Error function display
AN
: The system has a malfunction.
Blank display
BM
: Check if the battery has been loaded
correctly. Also check polarity (<+> and <->) of the battery.
Flat battery indication
BN
: If only the icon is shown on the
display, the battery should be replaced immediately.
10.Cleaning and disinfecting
Use an alcohol swab or cotton tissue moistened with alcohol (70%
Isopropyl) to clean the thermometer casing and the measuring
sensor. Ensure that no liquid enters the interior of the device.
Never use abrasive cleaning agents, thinners or benzene for
cleaning and never immerse the device in water or other cleaning
liquids. Take care not to scratch the surface of the sensor lens and
the display.
11.Battery Replacement
This device is supplied with 2 new, long-life 1.5V, size AAA
batteries. Batteries need replacing when this icon
BN
is the only
symbol shown on the display.
Remove the battery cover
BQ
by sliding it in the direction shown.
Replace the batteries – ensure correct polarity as shown by the
symbols in the compartment.
12.Guarantee
This device is covered by a
5
year guarantee
from the date of
purchase. During this guarantee period, at our discretion, Microlife
will repair or replace the defective product free of charge.
Opening or altering the device invalidates the guarantee.
The following items are excluded from the guarantee:
Transport costs and risks of transport.
Damage caused by incorrect application or non-compliance
with the instructions for use.
Damage caused by leaking batteries.
Damage caused by accident or misuse.
Packaging/storage material and instructions for use.
Regular checks and maintenance (calibration).
Accessories and wearing parts: Battery.
Should guarantee service be required, please contact the dealer
from where the product was purchased, or your local Microlife
service. You may contact your local Microlife service through our
website:
www.microlife.com/support
Compensation is limited to the value of the product. The guarantee
will be granted if the complete product is returned with the original
invoice. Repair or replacement within guarantee does not prolong
or renew the guarantee period. The legal claims and rights of
consumers are not limited by this guarantee.
13.Technical Specifications
Batteries and electronic devices must be disposed of in
accordance with the locally applicable regulations, not with
domestic waste.
Type:
Adjusted mode Ear Thermometer IR 310
Mode type:
Adjust mode
Measuring site:
Ear
Summary of Contents for IR310
Page 52: ...50 1 ACCUsens good 30 10 37 5 C 2 Microlife 15 3 3 m...
Page 64: ...62...