Which batteries and which procedure?
Use 4 new, long-life 1.5 V, size AA alkaline batteries.
Do not use batteries beyond their date of expiry.
Remove batteries if the device is not going to be used for a
prolonged period.
Using rechargeable batteries
You can also operate this device using rechargeable batteries.
Only use »NiMH« type reusable batteries.
Batteries must be removed and recharged when the flat
battery symbol appears. They should not remain inside the
device as they may become damaged (total discharge as a
result of low use of the device, even when switched off).
Always remove the rechargeable batteries if you do not
intend to use the device for a week or more.
Batteries cannot be charged in the blood pressure monitor.
Recharge batteries in an external charger and observe the
information regarding charging, care and durability.
7. Using a mains adapter
You can operate this device using the Microlife mains adapter (DC
6V, 600 mA).
Only use the Microlife mains adapter available as an orig
-
inal accessory appropriate for your supply voltage.
Ensure that neither the mains adapter nor the cable are
damaged.
1. Plug the adapter cable into the mains adapter socket
AT
in the
blood pressure monitor.
2. Plug the adapter plug into the wall socket.
When the mains adapter is connected, no battery current is
consumed.
8. Bluetooth
®
Function
This device can be used in conjunction with a smartphone running
the »Microlife Connected « App.
Bluetooth
®
can be activated manually by pressing and holding the
ON/OFF button
1
for about 4 seconds until the Bluetooth symbol
begins flashing
AO
.
Bluetooth
®
is automatically activated
AO
after a measurement has
been completed.
Bluetooth pairing and set up is required to use this device
with the »Microlife Connected « App. It is recom
-
mended to manually activate Bluetooth
®
and connect to the
»Microlife Connected « App to synchronize the date
and time settings between the device and the smartphone.
In case of difficulties connecting the device with the »Micro
-
life Connected « App, try to reset the Bluetooth
®
connection of this device to default settings:
Press and hold the ON/OFF button
1
for 8 seconds until
»
CL Pr
« is displayed. This clears the Bluetooth
®
pairing
records on the device and reset it to factory settings. Blue
-
tooth
®
automatically activates after the reset.
For more detailed information visit www.microlife.com/connect.
9. PC-Link functions
This device can be used in conjunction with a personal computer
(PC) running the Microlife Blood Pressure A (BPA+) soft
-
ware. The memory data can be transferred to the PC by
connecting the monitor via a cable.
If no download-voucher and cable is included download the BPA+
software from www.microlife.com/software and use a Micro-USB
cable.
10.Error messages
If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g.
«Err 3»
, is displayed.
Error
Description Potential cause and remedy
«Err 1»
BK
Signal too
weak
The pulse signals on the cuff are too
weak. Re-position the cuff and repeat
the measurement.*
«Err 2»
BT
-B
Error signal During the measurement, error signals
were detected by the cuff, caused for
instance by movement or muscle
tension. Repeat the measurement,
keeping your arm still.
«Err 3»
BT
-C
Abnormal
cuff pressure
An adequate pressure cannot be gener
-
ated in the cuff. A leak may have
occurred. Check that the cuff is correctly
connected and is not too loose. Replace
the batteries if necessary. Repeat the
measurement.