background image

Instruction Manual

EN

3

BP 3SZ1-1

Watch

BP

O3

Professional 24-hour

Ambulatory Blood

Pressure Monitor

Summary of Contents for BP 3SZ1-1

Page 1: ...Instruction Manual EN 3 BP 3SZ1 1 WatchBP O3 Professional 24 hour Ambulatory Blood Pressure Monitor...

Page 2: ...and ISO 81060 2 2013 protocol equivalence Function for Options AFIB Version The device can detect atrial fibrillation if present during blood pressure measurement CBP Version The device can determine...

Page 3: ...eparing the measurement arm 15 Positioning the cuff and device 15 Taking blood pressure measurement Next measurement indicator 16 Measurement reminder 16 During blood pressure measurements 17 Initiati...

Page 4: ...rmat for data analysis can be generated Device Version The device is upgradable for obtaining special features There are three different versions of the device Standard version standard ABPM AFIB vers...

Page 5: ...O3 Cuff Size L 4 Carrying Pouch 5 Data Cable 6 AAA Alkaline Batteries x 4 7 Tube holder x 3 8 100 cm cuff tube x 2 9 Instruction Manual 10 Patient diary 11 Quick start guide Download the latest Watch...

Page 6: ...6 Product Overview Battery compartment SYS mmHg DIA mmHg PUL min START STOP Start Stop button On Off button USB port Medication button Cuff connector Product description...

Page 7: ...7 WatchBP O3 Bluetooth indicator Battery indicator AF indicator Count down indicator Medication indicator cPP indicator Heart indicator Time SYS cSBP DIA cDIA Countdown min to next measurement Pulse C...

Page 8: ...ear of the device and insert 4x AAA alkaline batteries according to the battery polarities and close the compartment Press the On Off button to activate the device The device displays sequentially the...

Page 9: ...en successfully connected with WatchBP Analyzer PC software If the device and WacthBP Analyzer software is connected successfully USb is displayed on the LCD screen of the device The device ID model v...

Page 10: ...ween 00 to 23 and the asleep Night measurement period can be set to any hour between 00 to 23 Measurement time intervals can be set to 5 10 15 20 30 or 60 minutes for both awake and asleep time period...

Page 11: ...o show the measured blood pressure values on the LCD screen and to provide a beep before each measurement tick the squares to the left The display of blood pressure values on the LCD screen can be ena...

Page 12: ...available in both nylon and cotton S M L XL S XL size cuffs can be ordered optionally M L size cuffs are supplied as standard with the device Connecting the cuff tube to the device Push the end of th...

Page 13: ...and should be mounted over the patient s shoulder to the other site of the body Adjust the tube holder on the belt of the pouch so it is properly positioned on the shoulder Add the pouch to the belt t...

Page 14: ...a short series of beeps to notify the patient of the coming measurement optionally The beep is switched off by default and can be enabled via software setting See instruction manual of the WatchBP An...

Page 15: ...ally repeat the measurement after a two minute countdown should an error occur during measurement At any time a single measurement can be stopped by pressing the Start Stop or ON OFF button If the rep...

Page 16: ...he arm Readings are displayed at the end of the manually initiated blood pressure measurement When a central blood pressure measurement is performed The LCD screen switches every 2 seconds after the m...

Page 17: ...illation is palpitations However many people have no symptoms and therefore may remain undetected whereas diagnosing atrial fibrillation early followed by adequate treatment can largely reduce the cha...

Page 18: ...alue are then determined1 The time that is needed to determine the central blood pressure value may vary among patients i e with faster heart rate less time is required for collecting the number Sung...

Page 19: ...e patient can record the time of medication intake by pressing and holding the medication button for 2 seconds The pill icon flashes twice and displays for 1 min before the LCD screen returns to displ...

Page 20: ...and Central Blood Pressure measurement of the device can be activated through the WatchBP Analyzer A 16 digit activation key is needed specifically to match the device ID for activation Please contact...

Page 21: ...ad and follow the procedure of the WatchBP Analyzer to transfer the measurement data to a computer Deleting measurements The measurement data on the device will be automatically deleted after clicking...

Page 22: ...ed The battery icon and voltage number are also displayed on the LCD screen of the device How to replace batteries Open the battery compartment at the back of the device Replace the batteries ensure c...

Page 23: ...device Press and hold the Start Stop button for 7 seconds the Bluetooth icon flashes and starts pairing mode The unique 6 digit device ID of the unit is displayed Connect the device and confirm pairin...

Page 24: ...ons in the individual sections of the instruction manual Do not connect the device to a computer until prompted to do so by the computer software Follow the Instructions for Use This document provides...

Page 25: ...solution 1 10 dilution of household chloride bleach in water Isopropyl alcohol 70 solution Then wipe the exterior of the device with a soft dry cloth Cleaning the cuff Take out the bladder Fold and pl...

Page 26: ...be tested for accuracy every 2 years or after mechanical impact e g Being dropped Please contact Microlife to arrange an accuracy test Disposal Batteries and electronic instruments must be disposed o...

Page 27: ...blem occurs repeatedly If you think the results are unusual please read through the information in this instruction manual carefully Error Description Potential cause and remedy Er 1 Signal too weak T...

Page 28: ...ugh the checklist for performing reliable measurements and then repeat the measurement Er 11 Signal too weak during central blood pressure measurement The pulse signals on the cuff are too weak Re pos...

Page 29: ...esult can therefore be displayed Read through the checklist for performing reliable measurements and then repeat the measurement Er F The device has gone into single fault condition Single fault condi...

Page 30: ...o high The pressure in the cuff is too high over 299 mmHg OR the pulse is too high over 239 beats per minute Relax for 5 minutes and repeat the measurement LO Pulse too low The pulse is too low less t...

Page 31: ...inute Cuff pressure display Range 0 299 mmHg Resolution 1 mmHg Static accuracy pressure within 3 mmHg Pulse accuracy 5 of the readout value Power source 4X1 5 V Batteries size AAA Expected service lif...

Page 32: ...32...

Page 33: ...ntee is valid only on presentation of the guarantee card completed by the owner confirming date of purchase or purchase receipt Batteries and wearing parts are not covered by this guarantee Product Wa...

Page 34: ...watchbp com Asia Microlife Corporation 9F No 431 Ruiguang Rd Neihu Dist Taipei City 114 Taiwan R O C Tel 886 2 8797 1288 Fax 886 2 8797 1283 Email watchbp microlife com tw www watchbp com North Centr...

Page 35: ...EN 35 WatchBP O3...

Page 36: ...36...

Reviews: