c. Verify that the process fluid is not flashing or condensing, and that it does not
contain particles that can settle out.
d. Remove or reduce sources of electromechanical noise if appropriate.
2.
Start the zero process.
a. Navigate to the Zero menu:
OFFLINE MAINT
>
OFFLINE CALIB
>
VEL ZERO
.
b. Scroll to
CAL VEL/ZERO
and activate
Select
, then select
CAL/YES?
.
The status LED changes to flashing yellow while the zero is in progress.
3.
Read the zero result on the display.
The display reports
CAL PASS
if the zero was successful, or
CAL FAIL
if it was not.
Postrequisites
Restore normal flow through the sensor by opening the valves.
Need help?
If the zero fails:
•
Ensure that there is no flow through the sensor, then retry.
•
Remove or reduce sources of electromechanical noise, then retry.
•
Set
Zero Time
to a lower value, then retry.
•
If the zero continues to fail, contact Micro Motion.
•
If you want to return the meter to operation using a previous zero value, choose
OFFLINE
MAINT
>
CAL
>
RESTORE VEL/ZERO
.
2.4.2
Zero the meter using ProLink III
Zeroing the meter establishes a baseline for the velocity indicator by analyzing the sensor's
output when there is no flow through the sensor tubes.
Important
In most cases, the default value (0) is adequate for flow indication. Do not zero the meter unless the
velocity switch is being triggered incorrectly and you need to apply an offset to the velocity
measurement.
Prerequisites
Important
Do not zero the meter if a high-severity alert is active. Correct the problem, then zero the meter. You
may zero the meter if a low-severity alert is active.
Procedure
1.
Prepare the meter:
a. Stop flow through the sensor by shutting the downstream valve, and then the
upstream valve if available.
Quick start
10
Micro Motion
®
Compact Density Meters (CDM)
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...