Page 98
Appendix | Process Interface for thermoIMAGER TIM 40
thermoMETER TIM 8 / thermoIMAGER TIM 40
A 12
Process Interface for thermoIMAGER TIM 40
A 12.1
Analog Output
Fig. 58 Analog output
For voltage measurements the minimum load impedance should be 10 KOhm.
The analog output can be used as a digital output. The voltage for
no alarm
and
alarm on
can be set
within the software. The analog output (0 … 10 V) has a 100 Ohm resistor in raw. With a maximum current of
10 mA the voltage drop is 1 V.
Having an alarm LED with a forward voltage of 2 V the analog output value for
alarm on
should be 3 V as
maximum.