background image

Operating Instructions

opto

NCDT

 1900-IE

EtherCAT

EtherNet/IP

Profinet

Intended Use

The optoNCDT 1900 is designed for use in industrial and laboratory applications. It is used for 
measuring displacement, distance and position as well as in quality control and dimensional 
testing.
The sensor must only be operated within the limits specified in the technical data, see operating 
instructions, Chap 3.3. The sensor must be used in such a way that no persons are endangered 
or machines and other material goods are damaged in the event of malfunction or total failure of 
the sensor. Take additional precautions for safety and damage prevention in case of safety-relat-
ed applications. 

Warnings

Connect the power supply according to the safety regulations for electrical equipment. The sup-
ply voltage must not exceed the specified limits.

 

> Risk of injury, damage to or destruction of the sensor.

Avoid constant exposure of the sensor to splashes of water. Avoid exposure of sensor to aggres-
sive media (detergents, cooling emulsions).

 

> Damage to or destruction of the sensor

Avoid shock and vibration to the sensor. Protect the sensor cable against damage.

 

> Damage to or destruction of the sensor, failure of the measuring device.

Proper Environment

 - Protection class: IP67 (applies only when sensor cable is plugged in)

Lenses are excluded from the protection class. Contamination of the lenses causes impairment 
or failure of the function.

 - Temperature range

 - Humidity:

5 ... 95% (non-condensing)

 

ƒ

Operation:  0 ... +50 °C

 - Ambient pressure: Atmospheric pressure

 

ƒ

Storage:

-20 ... +70 °C

Sensor Mounting

The optoNCDT 1900 sensor is an optical system for measurements with micrometer accuracy. 
Ensure careful handling during installation and operation!

  Mount the sensor only to the existing through-bores (mounting holes) on a flat surface. Any 

type of clamping is not permitted.

  Mount the sensor using two M4 screws or the via the through-bores for M3 using the 

screws from the accessories. 

Measuring Range, Start of Measuring Range

SMR

MR

Digital value

 1

262077

98232 (SMR)

131000 (MMR)

163768 (EMR)
262078

Mounting with diffuse reflection. If the laser beam does 
not strike the object surface at a perpendicular angle, 
measurements might be inaccurate.

MR  = Measuring range
SMR = Start of measuring range
MMR = Mid of measuring range
EMR = End of measuring range

1) For displacement values without zero setting or 

mastering.

Mounting

Bolt connection

Direct fastening

0.75 Nm

2.0 Nm

min 10

M3 x 40; ISO 4762, A2-70

M4; ISO 4762, A2-70;  
Screwing depth min. 10 mm

optoNCDT

laser

aperture

Laser-

austritts-

öffnung

LASER RADIATION

AVOID DIRECT EYE EXPOSURE

CLASS 3R LASER PRODUCT

IEC 60825-1: 2014

P 5 mW;

 

= 658 nm

COMPLIES WITH 21 CFR 1040.10 AND 1040.11

EXCEPT FOR CONFORMANCE WITH IEC 60825-1

ED. 3., AS DESCRIBED IN

LASER NOTICE NO. 56, DATED MAY 8, 2019.

Drawings, Free Space 

 32 

 18 

 6

 6 

 14 

 31 (1.22) 

 4x M4 

 45 (1.77) 

 4 (.16) 

 70 (2.76)

 5 (.20) 

 2 (.08) 

 

 2x ø6 H7 

+0.012

0

 2x 6 H7 

+0.012

0

 6 (.24) 

 60 (2.36) 

 SMR 

 

MR 

 

 2 (.08) 

Keep this area free from other 

Limits for free space

light sources and/or their reflections.
Dimensions in mm (inches)

 (1.26)

 (.71) 

 (.55)

(.24)

 (.24)

 11 

(.43) 

MR

2/2LL 6/6LL 10/10LL 25/25LL 50/50LL 100

200

500

SMR

15

17

20

25

40

50

60

100

X

23

27

33

33

36

37

39

43

Y

3

9

14

33

45

75

130

215

Dimensions in mm

Dimensional drawing, drilling 
pattern, mounting plate

Alignment by  
centering elements (optional)

ILD1900-IE with pigtail

Pin assignment

Signal

Wire color  
PC1900-IE-x/OE-RJ45 

Comments

V

+

Red

Power supply

11 ... 30 VDC,  
typ. 24 VDC

GND

Blue

Reference 
ground

Laser on/off +

Black

Switching input

Laser in the sensor is active if both 
pins are connected to each other.

Laser on/off -

Violet

Open end connections, PC1900-IE-x/OE-RJ45

Signal

Pin Comments

2

1

8

7

V

+

1

Power supply

11 ... 30 VDC,  
typ. 24 VDC

GND

2

Reference ground

Laser on/off +

7

Switching inputs

Laser on/off -

8

Pigtail connections on the sensor, 12-pin connector, M12, pin side pigtail cable connector

Supply voltage

Nominal value: 24 V DC (11 ... 30 V, P < 3 W).

Industrial Ethernet with PoE

Sensor supply is via a PoE-capable switch. Phantom powering (PoE) is possible via the 

 - PC1900-IE-x/RJ45 
 - PC1900-IE-x/OE-RJ45

cables.

Industrial Ethernet without PoE

Sensor supply is via the PC1900-IE-x/OE-RJ45 cable.

ILD1900-IE

1

2

11 ... 
30 VDC

Sensor 

pin

PC1900-IE-

x

/OE-RJ45 

Color

Power supply

1

Red

V

+

2

Blue

GND

As an alternative to PoE, the sensor can be supplied with the optional PS2020 power supply unit.
Voltage supply only for measuring devices, not to be used for drives or similar sources of impulse 
interference at the same time. MICRO-EPSILON recommends using an optional available power 
supply unit PS2020 for the sensor. 

  Only turn on the power supply after wiring has been completed.
  Connect the inputs Pin 1 and Pin 2 at the sensor with a 24V power supply.

Electrical Connections

RJ45 Connection, PoE, Laser On/Off via Software

PC1900-IE-

x

/RJ45

PC / Ethernet

PoE Switch

PLC

RJ45 Connection, PoE, Laser On/Off via Hardware

PC1900-IE-

x

/OE-RJ45

Laser On/Off

PC / Ethernet

PoE Switch

PLC

RJ45 Connection, Supply via Optional Power Supply Unit, Laser On/Off via Hardware

PS 2020

PC1900-IE-

x

/OE-RJ45

Laser On/Off

PC / Ethernet

Switch

PLC

X9771444.01-A022072MSC

MICRO-EPSILON MESSTECHNIK GmbH & Co. KG
Königbacher Str. 15 

 94496 Ortenburg / Deutschland

Tel. +49 (0) 8542 / 168-0 

 Fax +49 (0) 8542 / 168-90

[email protected] 

 www.micro-epsilon.com 

Your local contact:

 

www.micro-epsilon.com/contact/worldwide/

You can find more information about the sensor in the operating instructions. You will find them 
online at:  

www.micro-epsilon.de/displacement-position-sensors/laser-sensor/optoNCDT_1900/

ø4.8 (.19)

ø6 H7 

+0.012

0

1.6 

±0.05

 

60  (2.36) 

±0.1

0.02

A

0.02

A

optoNCDT

0.3 m

 47.5
(1.87)

 14

(.55) 

M12 

Laser Safety

The optoNCDT 1900 operates with a semiconductor laser with a wavelength of 670 nm (visible/
red) or 658 nm (visible/red). Operation of the laser is indicated visually by the LED 

state

 on the 

sensor. When operating the optoNCDT 1900 sensors, the relevant regulations according to IEC 
60825, Part 1 of 05/2014 and the applicable accident prevention regulations must be followed.

i

  If both warning labels are covered over when the unit is installed, the user must ensure that 

supplementary labels are applied.

Laser Class 2

The sensors fall within laser class 2. The laser is operated on a pulsed mode, the maximum optical 

power is ≤ 1 mW. The pulse frequency depends on the adjusted measuring rate (0.25 ... 10 kHz). 

The pulse duration of the peaks is regulated depending on the measuring rate and reflectivity of 
the target and can be 4 up to 3995 μs.

LASER RADIATION

DO NOT STARE INTO BEAM

CLASS 2 LASER PRODUCT

IEC 60825-1: 2014

P 1mW; 

=670nm

COMPLIES WITH 21 CFR 1040.10 AND 1040.11

 EXCEPT FOR CONFORMANCE WITH IEC 60825-1

 ED. 3., AS DESCRIBED IN

LASER NOTICE NO. 56, DATED MAY 8, 2019.

 

Laser radiation. Close your eyes or 
immediately turn away if the laser 
beam hits the eye. Irritation or injury 
of the eyes possible.

Laser warning sign on 
the sensor housing

Laser labels on the sensor cable

Laser Class 3R

The sensors fall within laser class 3R. The laser is operated on a pulsed mode, the maximum 

optical power is ≤ 5 mW. The pulse frequency depends on the adjusted measuring rate (0.25 ... 10 

kHz). The pulse duration of the peaks is regulated depending on the measuring rate and reflectivity 
of the target and can be 4 up to 3995 μs.

laser

aperture

Laser-

austritts-

öffnung

Laser radiation. Use suitable pro-
tective equipment and close your 
eyes or immediately turn away if the 
laser beam hits the eye. Irritation or 
injury of the eyes possible.

Laser warning sign on 
the sensor housing

Laser labels on the sensor cable

i

 Observe the laser protection regulations.

Accordingly, the following applies: The accessible laser radiation is harmful to the eyes. Looking 
directly into the laser beam is harmful to the eyes with laser class 3R devices. Reflections of shiny 
or mirroring surfaces are also harmful to the eyes.
Class 3R laser sensors require a laser protection officer. Mark the laser area recognizable and 
everlasting. During operation the laser area has to be restricted and marked.

Sensor with laser labels