Operating Instructions
opto
NCDT
1900-IE
EtherCAT
EtherNet/IP
Profinet
Intended Use
The optoNCDT 1900 is designed for use in industrial and laboratory applications. It is used for
measuring displacement, distance and position as well as in quality control and dimensional
testing.
The sensor must only be operated within the limits specified in the technical data, see operating
instructions, Chap 3.3. The sensor must be used in such a way that no persons are endangered
or machines and other material goods are damaged in the event of malfunction or total failure of
the sensor. Take additional precautions for safety and damage prevention in case of safety-relat-
ed applications.
Warnings
Connect the power supply according to the safety regulations for electrical equipment. The sup-
ply voltage must not exceed the specified limits.
> Risk of injury, damage to or destruction of the sensor.
Avoid constant exposure of the sensor to splashes of water. Avoid exposure of sensor to aggres-
sive media (detergents, cooling emulsions).
> Damage to or destruction of the sensor
Avoid shock and vibration to the sensor. Protect the sensor cable against damage.
> Damage to or destruction of the sensor, failure of the measuring device.
Proper Environment
- Protection class: IP67 (applies only when sensor cable is plugged in)
Lenses are excluded from the protection class. Contamination of the lenses causes impairment
or failure of the function.
- Temperature range
- Humidity:
5 ... 95% (non-condensing)
Operation: 0 ... +50 °C
- Ambient pressure: Atmospheric pressure
Storage:
-20 ... +70 °C
Sensor Mounting
The optoNCDT 1900 sensor is an optical system for measurements with micrometer accuracy.
Ensure careful handling during installation and operation!
Mount the sensor only to the existing through-bores (mounting holes) on a flat surface. Any
type of clamping is not permitted.
Mount the sensor using two M4 screws or the via the through-bores for M3 using the
screws from the accessories.
Measuring Range, Start of Measuring Range
SMR
MR
Digital value
1
262077
98232 (SMR)
131000 (MMR)
163768 (EMR)
262078
Mounting with diffuse reflection. If the laser beam does
not strike the object surface at a perpendicular angle,
measurements might be inaccurate.
MR = Measuring range
SMR = Start of measuring range
MMR = Mid of measuring range
EMR = End of measuring range
1) For displacement values without zero setting or
mastering.
Mounting
Bolt connection
Direct fastening
0.75 Nm
2.0 Nm
min 10
M3 x 40; ISO 4762, A2-70
M4; ISO 4762, A2-70;
Screwing depth min. 10 mm
optoNCDT
laser
aperture
Laser-
austritts-
öffnung
LASER RADIATION
AVOID DIRECT EYE EXPOSURE
CLASS 3R LASER PRODUCT
IEC 60825-1: 2014
P 5 mW;
= 658 nm
COMPLIES WITH 21 CFR 1040.10 AND 1040.11
EXCEPT FOR CONFORMANCE WITH IEC 60825-1
ED. 3., AS DESCRIBED IN
LASER NOTICE NO. 56, DATED MAY 8, 2019.
Drawings, Free Space
32
18
6
6
14
31 (1.22)
4x M4
45 (1.77)
4 (.16)
70 (2.76)
5 (.20)
2 (.08)
Y
2x ø6 H7
+0.012
0
2x 6 H7
+0.012
0
6 (.24)
60 (2.36)
SMR
MR
X
2 (.08)
Keep this area free from other
Limits for free space
light sources and/or their reflections.
Dimensions in mm (inches)
(1.26)
(.71)
(.55)
(.24)
(.24)
11
(.43)
MR
2/2LL 6/6LL 10/10LL 25/25LL 50/50LL 100
200
500
SMR
15
17
20
25
40
50
60
100
X
23
27
33
33
36
37
39
43
Y
3
9
14
33
45
75
130
215
Dimensions in mm
Dimensional drawing, drilling
pattern, mounting plate
Alignment by
centering elements (optional)
ILD1900-IE with pigtail
Pin assignment
Signal
Wire color
PC1900-IE-x/OE-RJ45
Comments
V
+
Red
Power supply
11 ... 30 VDC,
typ. 24 VDC
GND
Blue
Reference
ground
Laser on/off +
Black
Switching input
Laser in the sensor is active if both
pins are connected to each other.
Laser on/off -
Violet
Open end connections, PC1900-IE-x/OE-RJ45
Signal
Pin Comments
2
1
8
7
V
+
1
Power supply
11 ... 30 VDC,
typ. 24 VDC
GND
2
Reference ground
Laser on/off +
7
Switching inputs
Laser on/off -
8
Pigtail connections on the sensor, 12-pin connector, M12, pin side pigtail cable connector
Supply voltage
Nominal value: 24 V DC (11 ... 30 V, P < 3 W).
Industrial Ethernet with PoE
Sensor supply is via a PoE-capable switch. Phantom powering (PoE) is possible via the
- PC1900-IE-x/RJ45
- PC1900-IE-x/OE-RJ45
cables.
Industrial Ethernet without PoE
Sensor supply is via the PC1900-IE-x/OE-RJ45 cable.
ILD1900-IE
1
2
11 ...
30 VDC
Sensor
pin
PC1900-IE-
x
/OE-RJ45
Color
Power supply
1
Red
V
+
2
Blue
GND
As an alternative to PoE, the sensor can be supplied with the optional PS2020 power supply unit.
Voltage supply only for measuring devices, not to be used for drives or similar sources of impulse
interference at the same time. MICRO-EPSILON recommends using an optional available power
supply unit PS2020 for the sensor.
Only turn on the power supply after wiring has been completed.
Connect the inputs Pin 1 and Pin 2 at the sensor with a 24V power supply.
Electrical Connections
RJ45 Connection, PoE, Laser On/Off via Software
PC1900-IE-
x
/RJ45
PC / Ethernet
PoE Switch
PLC
RJ45 Connection, PoE, Laser On/Off via Hardware
PC1900-IE-
x
/OE-RJ45
Laser On/Off
PC / Ethernet
PoE Switch
PLC
RJ45 Connection, Supply via Optional Power Supply Unit, Laser On/Off via Hardware
PS 2020
PC1900-IE-
x
/OE-RJ45
Laser On/Off
PC / Ethernet
Switch
PLC
X9771444.01-A022072MSC
MICRO-EPSILON MESSTECHNIK GmbH & Co. KG
Königbacher Str. 15
•
94496 Ortenburg / Deutschland
Tel. +49 (0) 8542 / 168-0
•
Fax +49 (0) 8542 / 168-90
•
www.micro-epsilon.com
Your local contact:
www.micro-epsilon.com/contact/worldwide/
You can find more information about the sensor in the operating instructions. You will find them
online at:
www.micro-epsilon.de/displacement-position-sensors/laser-sensor/optoNCDT_1900/
ø4.8 (.19)
ø6 H7
+0.012
0
1.6
±0.05
60 (2.36)
±0.1
0.02
A
0.02
A
optoNCDT
0.3 m
47.5
(1.87)
14
(.55)
M12
Laser Safety
The optoNCDT 1900 operates with a semiconductor laser with a wavelength of 670 nm (visible/
red) or 658 nm (visible/red). Operation of the laser is indicated visually by the LED
state
on the
sensor. When operating the optoNCDT 1900 sensors, the relevant regulations according to IEC
60825, Part 1 of 05/2014 and the applicable accident prevention regulations must be followed.
i
If both warning labels are covered over when the unit is installed, the user must ensure that
supplementary labels are applied.
Laser Class 2
The sensors fall within laser class 2. The laser is operated on a pulsed mode, the maximum optical
power is ≤ 1 mW. The pulse frequency depends on the adjusted measuring rate (0.25 ... 10 kHz).
The pulse duration of the peaks is regulated depending on the measuring rate and reflectivity of
the target and can be 4 up to 3995 μs.
LASER RADIATION
DO NOT STARE INTO BEAM
CLASS 2 LASER PRODUCT
IEC 60825-1: 2014
P 1mW;
=670nm
COMPLIES WITH 21 CFR 1040.10 AND 1040.11
EXCEPT FOR CONFORMANCE WITH IEC 60825-1
ED. 3., AS DESCRIBED IN
LASER NOTICE NO. 56, DATED MAY 8, 2019.
Laser radiation. Close your eyes or
immediately turn away if the laser
beam hits the eye. Irritation or injury
of the eyes possible.
Laser warning sign on
the sensor housing
Laser labels on the sensor cable
Laser Class 3R
The sensors fall within laser class 3R. The laser is operated on a pulsed mode, the maximum
optical power is ≤ 5 mW. The pulse frequency depends on the adjusted measuring rate (0.25 ... 10
kHz). The pulse duration of the peaks is regulated depending on the measuring rate and reflectivity
of the target and can be 4 up to 3995 μs.
laser
aperture
Laser-
austritts-
öffnung
Laser radiation. Use suitable pro-
tective equipment and close your
eyes or immediately turn away if the
laser beam hits the eye. Irritation or
injury of the eyes possible.
Laser warning sign on
the sensor housing
Laser labels on the sensor cable
i
Observe the laser protection regulations.
Accordingly, the following applies: The accessible laser radiation is harmful to the eyes. Looking
directly into the laser beam is harmful to the eyes with laser class 3R devices. Reflections of shiny
or mirroring surfaces are also harmful to the eyes.
Class 3R laser sensors require a laser protection officer. Mark the laser area recognizable and
everlasting. During operation the laser area has to be restricted and marked.
Sensor with laser labels