5.3.2.3 Channel Overview
Settings can be applied to each individual measuring channel in the Channel Overview submenu.
Name
Value
Description
Channel 1...28
You can switch between the available measuring channels via the
drop-down menu.
Activate output
The channel can be activated or deactivated by checking
Activate
output
.
Color value
Displays the color values depending on the selected color space
(XYZ, xyZ, Luv, uvL and RGB)
Dominant wavelength (nm):
Displays the dominant wavelength in nanometers.
Color temperature (Kelvin):
Displays the color temperature in Kelvin.
Timestamp (seconds):
Timestamp for the measuring channels since sensor start-up
Integration time (ms):
1, 2, 4, 8, 16,
32, 64, 128,
256, 512, 1024,
2048, 4096,
8192, 16384
You can set a value for the
Integration time
via the drop-down
menu.
By adjusting the sensitivity (modulation), the optimal integration
time for the lighting situation of the receiver can be determined. If
the selected setting does not appear to be optimal, the value for the
integration time can be adjusted via the drop-down menu.
Gain level:
1, 2, 4, 8, 16,
32, 64, 128,
256, 512, 1024,
2048
By adjusting the sensitivity (modulation), the optimal amplification
value for the lighting situation of the receiver can be determined. If it
does not appear to be optimal, it can be adjusted via the drop-down
menu.
Averaging:
Value
Choose the number of values to be used for forming a moving aver-
age.
Fields with gray background require a selection.
Value
Fields with dark border require entry of a
value.
By clicking on the button, a sensible combination of the parameters
Integration time
and
Gain
level
is determined automatically for the selected channel.
This button can be used to reset the intensity adjustment for the current channel.