Appendix | Optional Accessories
thermoMETER CSmicro
A 1.3
Air Purge Collars SF02/ SF15/ 2W/ 2WM-2
The lens must be kept clean at all times from dust, smoke, fumes and other contaminants in order to avoid
reading errors. These effects can be reduced by using an air purge collar.
i
Make sure to use oil-free, technically clean air, only!
Fig. 13 Standard air purge
collar; fits to the mounting
bracket; hose connection:
3x5 mm [TM-AP-CS]/ for
sensors with D:S ≥ 10:1
Fig. 14 Laminar air purge collar - the
sideward air outlet prevents a cooling
down of the object in short distances;
hose connection: 3x5 mm [TM-APL-CS]
Fig. 15 Combination of the
laminar air purge collar with the
bottom section of the mounting
fork
Dimensions in mm, not to scale
A combination of the laminar air purge collar with the bottom section of the mounting fork allows an adjust-
ment in two axis [TM-APL-CS +TM-MG-CS].
i
The needed amount of air (approximately 2 ... 10 l/ min.) depends on the application and the installation
conditions on-site.