21
probable order)
•
Ensure the device is in the Main Menu screen.
•
Electronics failure. Contact support.
Problem Fault
Symptoms:
•
Test begins automatically.
•
Volume accumulates automatically without
the subject blowing.
•
Very small VC or FVC test displayed.
Possible
Solutions: (In
probable order)
•
Device (or if using remote flowhead, the
flowhead) is not stationary at the start of test.
Hold it steady until the
‘Ready to Blow’ prompt
appears.
•
Return to previous menu and re-enter the test.
Problem Fault
Symptoms:
•
Cannot read screen
Possible
Solutions: (In
probable order)
•
Ensure the on/off button was pressed.
•
The batteries may be low, plug in USB cable
attached to a computer or replace the batteries.
•
Electronics failure. Contact Support.
Problem Fault
Symptoms:
•
Reversed or no volume measurements.
Possible
Solutions: (In
probable order)
•
Ensure flowhead is properly connected/seated
on device (if using the remote flowhead also
check the tubing is connected correctly).
•
Electronics failure. Contact support.
9.1 Software Check
Information about the device may be obtained from the About option
in the Configuration Menu. This information may be used when inquiries
are made to Micro Direct or a service agent.
9.2 Product Useful Life Checks
To ascertain whether the device has exceeded its useful life, Micro Direct
recommends checking the flowhead and the real time clock.
The flowhead may be checked with the daily calibration verification, to be
completed by the clinician/healthcare professional and during the periodic
inspection of the device. Reference section 5.4. Calibration Verification
for details on how to check the device flowhead.
The clock will reset if the 3V coin cell battery has depleted. This will be
evident as the Time-Date option will be displayed every time the device is
turned on until the battery is replaced and the time/date is set again.
The coin cell has an expected battery life of 15 years+ which exceeds the
devices product life; therefore, a service is recommended at 10 years or
when product useful life checks dictate it is required.