15
A calibration verification report may be printed or saved if required, see
section 5.3 Reporting
If the procedure was followed correctly and the error icon is showing, the
calibration verification should be repeated. If the error continues to show,
contact the manufacturer using the contact information in this document.
Note: To exit the Calibration Verification screen without performing a
check, press the forward key
to return to the Configuration Menu
screen. The calibration verification will not be logged to the device
memory in this case.
Note: A calibration verification is recommended daily, if the device has
been stored or transported, if the flowhead is dropped or replaced and if
EMI (Electromagnetic Interference) is suspected or possible.
5.5 Configuration Options
To access the Configuration menu, press the
icon on the Main Menu
screen.
There are four configuration options:
1. The Subject option
to configure:
a. Posture: set the posture for the session to sitting
or
standing .
b. Weight: turn on
to enter the subject
’s weight or off
if
not required.
c.
Population Group: turn on
to enter the subject
’s
population group or off
if not required.
2. The Device setting option
to configure:
a. The device
which includes:
•
Selecting between % Predicted
or Z-score
.
The parameter selected will them be displayed in
the results screen.
•
Audio: turn audio off
and put the Micro
Spirometer into silent mode.
•
User Passcode: use
to lock the device so that
any user will be prompted for a passcode on start
up.