13
To prevent the device from switching off, press the Select
Function key
within three minutes of the last press or when
the reading is first displayed.
Cleaning
The plastic adapters should be replaced on a monthly basis or
after 250 tests (whichever is sooner). Between tests the adapter
may be immersed in a cold disinfection solution or mild detergent
sterilization solution then
MUST
be thoroughly rinsed in water and
dried completely before reuse.
The device can be cleaned using non-alcohol wipes (s
ee page
14)
. Please be certain not to touch the surface of the sensor or
allow moisture to do the same.
Cleaning with products that contain alcohol may result in
permanent contamination of the sensor.
Servicing
If your unit requires servicing or repair, then please contact Micro
Direct, Inc.
The CO Sensor should be replaced every two years.
If the device starts
with ‘Er’ (Error) message, it should be
returned for service.
Email:
Tel:
800-588-3381 or 207-786-7808