Start Measure
Please notice that you can ask for a change of module (bath or globule)
Noticed changer le module in the screen here under
Measure cycle
The cycle described below is automatic:
1.
Resetting of the measuring chain to zero,
2.
Scraping of the bath surface, if the bath is used, and scrapping selected or not in the
main menu
3.
Module goes up
4.
The height gauge goes down until it detects the module (bath or globule)
5.
The height gauge rotate below the component
6.
The height gauge goes up until it touches the component.
7.
The receptacle moves up until detection by the height gauge.
8.
The height gauge rotates and goes home (up)
9.
Only when using globule module:
Tthe system waits for the alignment with the globule chosen (4, 2 or 1 mm) and
confirm alignment by pressing the four or eight arrows on the key pad or key board
before clicking ok in this window :