THM1176/TFM1186 Use
r’s Manual
v 2.0 r 1.0
– 02/20
Copyright © 2020 Metrolab Technology SA
www.metrolab.com
22
Functional
Block
Command(s)
Function
*STB?, *SRE
USBTMC-USB488
Read and enable bits in Status Byte
USBTMC-USB488
Request service from host
*ESR?, *ESE
Read and enable bits in Standard Event
Status Register
*OPC, *WAI
Detect and wait for operation complete
:STATus …
Read and enable bits in OPERation and
QUEStionable registers
:SYSTem:ERRor
Query error queue
SYSTem
USBTMC-USB488
Clear input and output buffers
USBTMC-USB488
Remote/local control
– ineffective since
THM1176/TFM1186 has no local controls
*RST
Perform reset
*TST?
Perform self-test (not supported)
*IDN?
Return Instrument ID
:SYSTem:VERSion Return SCPI version
:SYSTem:HELP … Provide command help
:SYSTem:SLEEp
Power down analog subsystem
MMEMory
:MMEMory …
Manipulate FLASH memory files
DIAGnostic
:DIAGnostic: …
Initiate firmware upgrade
5-3 IEEE 488.2 / SCPI STATUS REGISTERS
IEEE 488.2 compliant instruments have at least two registers: the Status Byte and
the Standard Event Status Register. SCPI adds the Operation Status Register,
Questionable Status Register and Error/Event Queue. The diagram below, taken
from the SCPI standard, provides a good summary. This section describes how