ARM#:SYNChro?
[DEFault|OPTions]
Default
NONe
Arguments
POSitive
A transition from low to high of the synchronization signal found
on the encoder connector is used to leave the arm level.
NEGative
A transition from high to low of the synchronization signal found
on the encoder connector is used to leave the arm level.
EITHer
Is a conjunction of both POSitive and NEGative conditions.
NONe
Disables the synchronization function to the benefit of the error
signal that might be found on some encoders.
Example
ARM:SYNC POS
Configures the transition detector to generate an event when a low
to high condition occurs on the synchronization signal.
:ARM#[:SEQuence][:LAYER]:EXTernal[:SLOPe]
3-7-5
Configures the transition detector of the external input trigger signal to generate
an arm-condition on a specific edge.
Syntax
ARM#:EXTernal
POSitive|NEGative|EITHer|DEFault
ARM#:EXTernal?
[DEFault|OPTions]
Default
POSitive
Arguments
POSitive
A transition from low to high of the synchronization signal found
on the encoder connector is used to leave the arm level.
NEGative
A transition from high to low of the synchronization signal found
on the encoder connector is used to leave the arm level.
EITHer
Is a conjunction of both
POSitive
and
NEGative
conditions.
Example
TRIG:EXT POS
Configures the transition detector to generate an event each time a
positive transition occurs on the external input trigger signal.
:CALCulate#:FLUX[:CUMulative]
3-7-6
Enables or disables the accumulation of partial integrals values.