5 Troubleshooting
919 IC Autosampler plus
■■■■■■■■
29
5 Troubleshooting
5.1
Problems and their solutions
Problem
Cause
Remedy
The peristaltic pump
is pumping too little.
Peristaltic pump – Contact
pressure too weak.
Correctly set the contact pressure.
Peristaltic pump – Pump
tubing defective.
Replace the pump tubing.