![Metrohm 940 Professional IC Vario TWO Manual Download Page 87](http://html1.mh-extra.com/html/metrohm/940-professional-ic-vario-two/940-professional-ic-vario-two_manual_1779626087.webp)
■■■■■■■■■■■■■■■■■■■■■■
7 Technical specifications
940 Professional IC Vario TWO (2.940.2100)
■■■■■■■■
79
Gradient capabil-
ity
Isocratic or gradient (extendable to quaternary)
Profile
Step, linear, convex and concave
Resolution
< 1 nL/min
7.10
Injection valve
Actuator switch-
ing time
typ. 100 ms
Maximum operat-
ing pressure
35 MPa (350 bar)
Material
PEEK
7.11
Detector
You can find the technical specifications for the detector in the manual for
the detector.
7.12
Sample degasser
Material
Fluoropolymer
Resistance to sol-
vents
No restriction (except PFC)
Time to establish
vacuum
< 60 s
7.13
Power connection
Required supply
voltage
100 - 240 V ± 10% (autosensing)
Required fre-
quency
50 - 60 Hz ± 3 Hz (autosensing)
Power consump-
tion
■
65 W for typical analysis application
■
25 W standby (conductivity detector to 40 °C)
Power supply unit
■
Up to 300 W maximum, electronically monitored
■
internal fuse 3.15 A