5.8 Servicing the high-pressure pump
■■■■■■■■■■■■■■■■■■■■■■
52
■■■■■■■■
940 Professional IC Vario TWO (2.940.2100)
5.8
Servicing the high-pressure pump
NOTE
You can find a video sequence for this task in the
Multimedia Guide IC
Maintenance
Figure 14
High-pressure pump – Parts
1
Pressure screw, short (6.2744.070)
Fastened to the outlet valve holder.
2
Outlet valve holder
3
Fastening screw
4
Pump head
5
Inlet valve holder
6
Pressure screw, short (6.2744.070)
Fastened to the inlet valve holder.
7
Eluent aspiration tubing connector
Consists of a coupling with a pressure screw.
8
Pressure screw, short (6.2744.070)
Fastened to the pump output.
Maintenance interval
The following parts of the high-pressure pump must be serviced at least
once per year:
■
Inlet valve (6.2824.170)
■
Outlet valve (6.2824.160)
■
Piston seal (6.2741.020)
■
Zirconium oxide piston (6.2824.070)
Maintenance tasks can also be carried out if the following problems occur: