3.18 Initial start-up
■■■■■■■■■■■■■■■■■■■■■■
36
■■■■■■■■
940 Professional IC Vario TWO (2.940.2100)
4 Deaerating the high-pressure pump
■
Push the end of the purge needle (6.2816.040) over the end of
the purge capillary on the purge valve.
■
Insert the syringe (6.2816.020) in the Luer connector of the purge
needle.
■
Open the purge valve using the rotary knob (approx.
½
turn).
■
Switch on the high-pressure pump in MagIC Net.
■
Use the syringe to aspirate eluent until there are no more air bub-
bles in the eluent aspiration tubing.
■
Switch off the high-pressure pump in MagIC Net.
■
Seal the purge valve using the rotary knob.
■
Remove the syringe from the purge needle.
■
Pull the purge needle out of the purge capillary.
5 Rinsing the instrument without columns
■
Rinse the instrument (without columns) with eluent for 10
minutes.