3 Installation
930 Compact IC Flex SeS (2.930.1400)
■■■■■■■■
39
3
Remove the label from the lid of the CO
2
adsorption cartridge CW.
This uncovers the small opening in the lid of the CO
2
adsorption car-
tridge CW, through which air is aspirated.
The CO
2
adsorption cartridge CW is now ready for installation.
NOTE
The new CO
2
adsorption cartridge CW (6.2837.100) works
without
attached H
2
O adsorption cartridge.
Installing the CO
2
adsorption cartridge CW
Accessories
■
Adsorption cartridge holder (6.2057.080)
■
Prepared CO
2
adsorption cartridge CW (6.2837.100)
CAUTION
The following preparatory steps absolutely must be carried out for CO
2
suppression to operate correctly.