![Metrohm 892 Professional Rancimat Manual Download Page 45](http://html1.mh-extra.com/html/metrohm/892-professional-rancimat/892-professional-rancimat_manual_1779604045.webp)
■■■■■■■■■■■■■■■■■■■■■■
4 Operation
892 Professional Rancimat
■■■■■■■■
37
NOTICE
Different methods with different temperatures can be selected for
block A and block B.
The gas flow for the two blocks
A
and
B
cannot be switched on
and off individually for each block but only collectively. If methods
with different gas flows are loaded on the two blocks, then the
value from the block in which the gas flow is switched on is used.
2 Starting the heater (StabNet)
■
In the
Workplace
program part, click on the
[Start]
button for
Heater
within the block A area.
■
If required, also switch on the heater for
Block B
.
The color of the temperature display on the 892 Professional Ranci-
mat changes to red during the heating phase.
The color of the temperature display on the 892 Professional Ranci-
mat changes to green once the setpoint temperature has been
reached.
Heating period to 120 °C: approx. 45 min.
Heating period to 200 °C: approx. 60 min.
NOTICE
If you wish to switch off the heating, click on the
[Stop]
button.
3 Enter sample identification (StabNet)
■
Enter the sample identifications
Ident
and
Info 1 - 3
for all sam-
ple positions used.
The entries for
Ident
and
Info 1 - 3
can be selected from the
Text
templates
.
NOTICE
The
Info 2
and
Info 3
info fields can be activated in the sub-
window
Properties - Sample data
in the
Method
program
part.