4.2 Calibration functions
■■■■■■■■■■■■■■■■■■■■■■
28
■■■■■■■■
892 Professional Rancimat
The cell constants can be either entered manually or determined automat-
ically by means of a defined standard solution, e.g. the conductivity stan-
dard 100 µS/cm (6.2324.010).
NOTICE
You will find detailed information regarding the
StabNet
computer
program in the Tutorial.
4.2.2
Determining the temperature correction
Temperature correction
indicates the deviation of the current sample
temperature from the temperature of the heating block and forms part of
the method as a parameter.
It can be determined automatically with the calibrated, external tempera-
ture sensor.